Factors influencing on the discharge coefficients of sonic nozzle

소닉노즐의 유출계수에 영향을 미치는 인자에 관한 연구

  • 유성연 (충남대학교 기계설계공학과) ;
  • 이상윤 (충남대학교 대학원 기계설계공학과) ;
  • 박경암 (한국표준과학연구원 유체유동그룹)
  • Published : 1996.12.01

Abstract

Accuracy of gas flow measurements using sonic nozzle and factors which influence on the discharge coefficients of sonic nozzle are investigated with high pressure gas flow standard measurement system. The gas flow measurement system comprises two compressors, storage tank, temperature control loop, sonic nozzle test section, weighing tank, gyroscopic scale and data acquisition system. The experiments are performed at various nozzle throat diameter and inlet pressure. Overall uncertainty of discharge coefficients is estimated to less than .+-.0.2% and most of experimental data fall into this range. Dependence of discharge coefficients on the Reynolds number is good agreement with those suggested in ISO document. The influence of swirl on the discharge coefficients becomes greater as the nozzle throat diameter is enlarged. The discharge coefficient of conical nozzle shows about 4.5% lower discharge coefficients than those of toroidal nozzle, but variation trend with Reynolds number is very similar each other and reproducibility of data is very good.

Keywords