Comparison of Microwave with Conventional Wet-Digestion Methods for the Element Analysis of Plant and Compost

식물체 및 퇴비 성분 분석을 위한 마이크로웨이브 분해법과 기존 습식 분해법의 비교

  • Nam, Jae-Jak (National Institute of Agricultural Science & Technology) ;
  • Cho, Nam-Jun (National Institute of Agricultural Science & Technology) ;
  • Jeong, Yee-Guen (National Institute of Agricultural Science & Technology) ;
  • Lee, Sang-Hak (Dep. of Chemistry, Kyung-pook National University)
  • Published : 1998.09.30

Abstract

To verify the suitability of microwave digestion method for analyzing macro nutrients and metals. K, Ca, Mg, Cu, Zn, and Mn contents in plant and composts were investigated comparing with conventional wet digestion method. Because it takes short time to digest metal samples by using microwave method compared with conventional wet digestion method. Digestion time for metal samples was four to six hours long for wet digestion, while one hour long for microwave digestion. Determinant coefficients of Cu, Mn, Zn, Ca, K, and Mg contents between two analyzing methods were 0.99, 0.99, 0.94. 0.97, 0.98, and 0.91, respectively. In the case of Ca and K, microwave digestion method analyzed their contents relatively low compared with wet digestion method. Meanwhile, microwave digestion method analyzed the contents for Cu, Ma, and Zn relatively high compared with wet digestion method. This study results in that microwave digestion method could be an alternative to replacing conventional wet digestion method.

마이크로파 시료 분해법이 식물체와 퇴비 중의 K, Ca, Mg, Cu, Zn, Mn등을 분석하는데 적합한가를 검정하기 위하여 기존의 습식분해법과 비교하였다. 분석시간은 기존의 습식분해법이 4~6시간 걸리던 것에 비하여 마이크로파 분해법에서는 1시간 이하로 단축할 수 있었으며, 두 분석법간의 각 성분별 분석값의 상관계수($R^2$)는 각각 K 0.98, Ca 0.97, Mg 0.91, Zn 0.94, Mn 0.99, Cu 0.99를 나타내었다. 칼륨과 칼슘의 경우 기존 습식분해법의 분석값이 더 높은 경향을 나타내었고 구리, 망간, 아연의 경우는 마이크로파 분해법의 분석값이 더 높은 경향을 나타내었으나 마그네슘의 경우 일정한 경향이 없었다. 그러나 칼슘의 경우를 제외한 그 외 성분들의 편기도(biases)는 10% 이내였다.

Keywords