수두증 환자 치료를 위한 프로그램 밸브 단락 장치의 이용

The Use of the Programmable Valve Shunt System in the Management of Patients with Hydrocephalus

안성태;유도성;조경석;김재건;허필우;김달수;강준기
Ahn, Seong-Tae;Yoo, Do-Sung;Cho, Kyoung-Sock;Kim, Jae-Keon;Huh, Pil-Woo;Kim, Dal-Soo;Kang, Joon-Ki

  • 발행 : 20020200

초록

Objective:The goal of this study was to clarify the value of the programmable valve shunt system to readjust the pressure noninvasively for the adequate cerebrospinal fluid(CSF) drainage. Methods:The authors analyzed a single-center retrospective study of 54 patients suffering from hydrocephalus of various causes, as aneurysm(21 patients), trauma(13 patients), normal-pressure hydrocephalus(NPH, 7 patients) and so on. In 51 cases a programmable valve was implanted at the first shunt implantation. In three cases the shunt was replaced to a programmable valve system. Results:In 33 patients(61%) of cases valve pressure adjustment was required at least once(total number of readjustment 54, mean 1.7, maximum 5) for the reason of underdrainage(36) or overdrainge(18). The range of reprogrammed valve pressure was 10 to 90mm$H_{2}O$(mean 11.1${\pm}$15.9mm$H_{2}O$), the radiological finding improved 48.6% of frontal horn index at the beginning of reprogramming to 41.3%, and the clinical symptom related with hydrocephalus improved in 29 patients(87.8%) of pressure adjustment. Shunt reprogramming was necessary in patients with congenital hydrocephalus(100%), aneurysm(55%), NPH(71%), trauma(50%);the programmable valve proved particularly beneficial for overdrainge as subdural hygroma. Conclusion:The programmable valve was useful for the correction of overdrainge or underdrainage by the easy control of valve pressure without any invasive procedure.

키워드

참고문헌

  1. Belliard H, Roux FX, Turak B:The codman Medos programmable shunt valve. Evaluation of 53 implantation in 50 patients. Neurochirurgie(Fr) 42:139-146, 1996
  2. Black PM, Hakim R, Bailey NO:The use of the Codman-Medos programmable Hakim valve in the management of patients with hydrocephalus:illustrative cases. Neurosurgery 34:1110-1113, 1994 https://doi.org/10.1227/00006123-199406000-00040
  3. Carpenter MB:Core text of Neuroanatomy, ed 4. Baltimore:Williams & Wilkins, 1991, pp9-10
  4. Hakim S:Hydraulic and mechanical mis-mathcing of valve shunts used in the treatment of hydrocephalus:the need for a servo-valve shunt. Dev Med Child Neurol 15:646-653, 1973 https://doi.org/10.1111/j.1469-8749.1973.tb05175.x
  5. Han YM, Yoo DS, Kim D S, Huh PW, Cho KS, Kang JK:A clinical analysis of the ventriculo-peritoneal shunt with programmable shunt device. J Korean Neurosurg Soc 28:75-81, 1999
  6. Kamiryo T, Fujii Y, Kusaka M, Kashiwagi S, Ito H:Intracranial pressure monitoring using a programmable pressure valve and a telemetric intracranial pressure sensor in a case of slit ventricle syndrome after multiple shunt revisions. Childs Nerv Sys 7:233-234, 1991 https://doi.org/10.1007/BF00249403
  7. Miyake H, Ohta T, Kajimoto Y, Nagao K:A new ventriculoperitoneal shunt with a telemetric intracranial pressure sensor:clinical experience in 94 patients with hydrocephalus. Neurosurgery 40:931-935, 1997 https://doi.org/10.1097/00006123-199705000-00009
  8. Nulsen FE, Spitz EB:Treatment of hydrocephalus by direct shunt from ventricle to jugular vein. Surg Forum 2:399-403, 1952
  9. Reinprecht A, Czech T, Dietrich W:Clinical experience with a new pressure-adjustable shunt valve. Acta Neurochir 134:119-124, 1995 https://doi.org/10.1007/BF01417677
  10. Yamashita Na, Kamiya K, Yamada K:Experience with a programmable system. J Neurosurg 91:26-31, 1999 https://doi.org/10.3171/jns.1999.91.1.0026
  11. Zemack G, Romner B:Seven years of clinical experience with the programmable Codman Hakim valve:A retrospective study of 583 patients. J Neurosurg 92:941-948, 2000 https://doi.org/10.3171/jns.2000.92.6.0941