Design of the Blood Pressure Measurement System Using the Inflatable Oscillometric Method

가압식 오실로메트릭 방법을 사용한 혈압측정 시스템의 설계

  • Published : 2003.08.01

Abstract

Blood Pressure is one of the most fundamental Parameters which reflects physical conditions medically and the blood pressure measurement system using oscillometric method is a Non-Invasive Blood Pressure measurement device by measuring arterial Pressure through a cuff. In this paper. we designed a inflatable wrist blood pressure system which measures blood Pressure during the stepping inflation in the wrist cuff. The hardware system consists of a main power unit, a bladder in cuff unit, signal detection units, signal Processing units. a wireless data transmission unit, and a data display unit. We evaluated the reliability of this system by comparing and analyzing systolic. diastolic blood Pressure, and heart rate with other commercial blood Pressure measurement devices. Characteristic ratio values used to determine systolic and diastolic blood Pressure using MAA(Maximum Amplitude Algorithm) were 0.436 and 0.671 respectively.

혈압은 사람의 건강상태를 알 수 있는 가장 기본적인 의학적 파라메터 중 하나이며, 오실로메트릭 방법을 사용한 혈압측정 시스템은 팔 동맥에서 커프를 통해 나타나는 압력을 측정함으로써 혈압수치를 비침습적으로 모니터링 하는 장치이다. 된 논문에서는 커프에 공기를 주입하면서 혈압을 측정하는 가압식 손목형 혈압측정 시스템을 설계하였다 설계한 혈압측정 시스템은 전원부와 압력을 가하는 공기주머니를 포함하는 커프, 신호검출부, 신호처리부, 무선 송수신부 그리고 디스플레이부로 구성된다. 설계한 자동 혈압측정 시스템은 기존에 일반적으로 사용되고 있는 자동혈압측정 시스템과의 수축기, 이완기 혈압 및 심박동수의 비교분석을 통해 신뢰성을 평가하였다. 수축기 혈압과 이완기 혈압을 결정하기 위해 MAA (Maximum Amplitude Algorithm)를 사용하였고, 적용한 특성비율은 0.436. 0.671이 었다.

Keywords

References

  1. 2001년도 고혈압 · 당뇨관리사업 지침 보건복지부
  2. Ann. Biomedical Eng. v.17 The Korotkoff sound G.M.Drzewiecki;J.Melbin;A.Noordergraaf https://doi.org/10.1007/BF02368055
  3. Physiologic Experimentale v.2 Pression et vitesse du sang E.J.Marey
  4. The Application of Signal Averaging to the NIBP Oscillometric Waveform Lee A. Taylor;Ronald G. Bennett;Thomas J. Dorsett
  5. The Effects of Pulse Rate, Artefact and Pulse Strength on Oscillometric Non-Invasive Blood Pressure Measurements J.N.Amoore;W.B.Geake;D.H.Scott
  6. A Strategy for Determination of Systolic, Mean and Diastolic Blood Pressures from Oscillometric Pulse Profiles v.27 JCBT Moraes;M.Cerulli
  7. Proc 20th Ann Conf. Eng Med Biol. v.9 Indirect blood pressure measurement Doppler ultrasonic kinetoarterio graphy Ware R.W.;Laenger C.J.
  8. Anesthesiology v.85 Evaluation of a new Oscillometric Blood Pressure Monitor by Cuff Inflation Method M.Nakajima;T.Komatsu;T.Kimura;Y.Fujiwara;Y.Shimada
  9. Anesthesiology v.87 Measurement of Non-Invasive Oscillometric Blood Pressure During Cuff Inflation, Rather than deflation N.Morioka;M.Ozaki;K.Ozaki;DI Sessler;T.Matsukawa;H.Suzuki https://doi.org/10.1097/00000542-199709001-00431