An Experimental Study on the Influential Factors of Flow Measurement with Coriolis Mass Flowmeter

코리올리스 질량유량계의 유량측정에 영향을 미치는 인자에 관한 실험적 연구

  • 임기원 (한국표준과학연구원 유체유동그룹) ;
  • 이완규 (한국표준과학연구원 강도평가그룹)
  • Published : 2003.12.01

Abstract

Coriolis mass flowmeter(CMF), which can measure the mass flow directly, is getting rapid attention for the industrial and custody transfer purpose. In order to study the characteristics and the applicability of CMF, it is tested with the national flow standard system. Two types of sensing tube, U-type and straight type, are employed in the test. Water, spindle oil and viscosity Standard Reference Material whose viscosities are 1, 20 and, 67 $\textrm{mm}^2$/s, respectively, are studied. It is shown that the linearity of CMF is getting deteriorated as the fluid viscosity increases, which is due to the zero drift and the relaxation time of the fluid. To test its applicability in the case of high pressured gas, it is calibrated using compressed air, It shows 1∼l.6 % deviations compared to the calibration results using water. It concludes that the fluid velocity in CMF should be lower than the sonic velocity. In addition, the effects of the vibration from the pipeline and pump on CMF as well as the long term stability are studied.

Keywords

References

  1. Brain, T. J. S., 1969, 'The True Mass Flow-meter,' Flow Measurement Memo No. 26, Flow Measurement Division, NEL, Scotland
  2. Li, Y. T. and Lee, S. Y., 1953, 'A Fast Response True Mass Rate Flowmeter,' Trans. Am. Soc. Mech. Engrs, Vol. 75, pp835-841
  3. Keita, N. M., 1989, 'Contribution to the understanding of Zero Shift Effects in Coriolis Mass Flowmeters,' Flow Measurement and Instrumentation, Vol. 1, No. 1, pp 39-43, 1989 https://doi.org/10.1016/0955-5986(89)90008-3
  4. Furness, R., 1989, 'The Application, Standardi-sation and Future Use of Coriolis Type Mass Meters in the Oil and Process Industries,' Proc. International Conf. on Mass Flow Measurement, London
  5. Sultan, G. and Hemp, J., 1989, 'Modelling of the Coriolis Mass Flowmeter,' J. of Sound and Vibration, Vol. 132, No. 2, pp473-489 https://doi.org/10.1016/0022-460X(89)90640-8
  6. ISO 4185, 1983, Measurement of Liquid Flow in closed conduits - Weighing Method
  7. Lim, K. W., 2001, 'A Study on the Measure-ment Uncertainty of Flowmeter Calibrator,' J. of KSME (B), Vol.25, No.4, pp. P561-571
  8. Lim, K. W., 2002, 'Calibration Procedure of Spindle Oil Flow Standard System,'Doc. No.C-10-3-0101-2002(E), KRISS
  9. Park, K. A. and Lee, S. H., 2002, 'Calibration Procedure of High Pressure Gas Flow Standard System-Critical Flow Nozzle, Doc. No.C-10-4-0100-2002(E), KRISS
  10. Doc. No.C-10-4-0100-2002(E) Calibration Procedure of High Pressure Gas Flow Standard System-Critical Flow Nozzle Park,K.A.;Lee,S.H.