A Survey on Radiation Exposure of Patient in Mammography

유방 X선촬영 시 피폭선량에 대한 조사 연구

  • Kim, Hyung-Chul (Dept. of Diagnostic Radiology, Korea University Ansan Medical Center) ;
  • Cho, Pyung-Gon (Dept. of Diagnostic Radiology, Korea University Anam Medical Center) ;
  • Kim, Sung-Soo (Dept. of Diagnostic Radiology, Singu College) ;
  • Choi, Jong-Hak (Dept. of Radiologic Technology, College of Health Sciences, Korea University) ;
  • Kim, You-Hyun (Dept. of Radiologic Technology, College of Health Sciences, Korea University)
  • 김형철 (고려대학교 안산병원 진단방사선과) ;
  • 조평곤 (고려대학교 안암병원 진단방사선과) ;
  • 김성수 (신구대학) ;
  • 최종학 (고려대학교 보건대학 방사선과) ;
  • 김유현 (고려대학교 보건대학 방사선과)
  • Published : 2004.12.31

Abstract

The purpose of this study was to examine both patient exposure dose during mammography and the utility status of mammograpy equipments. The data of this study were collected through questionnaire survey for 278 medical facilities registered at Korean Hospital Association and finally 161 medical facilities's data were analyzed. According to data analysis, medical facilities of 14.9% used the average glandular dose of less than 0.5 mGy, $0.51{\sim}1.0\;mGy$ 8.6%, $1.01{\sim}1.5\;mGy$ 14.9%, $1.51{\sim}2.0\;mGy$ 11.1%, $2.01{\sim}2.5\;mGy$ 9.8%, $2.51{\sim}3.0\;mGy$ 33.3%, and 7.4% more than 3.01 mGy. It was found that medical facilities of 92.6% used less than 3 mGy, showing that this figure is similar to the limit value of 3 mGy recommended by Korea Food & Drug Administration(KFDA). Recently, international organizations such as ICRP associated with radiation protection suggests that less than 3 mGy of average mammary gland dose be used during mammography in case of using Mo target+Mo filter, film/screen system and craniocaudal projection with the breast pressed to 4.2 cm. The standard dose is being strictly observed and that of the limits is going down to 2 mGy or 1.5 mGy. The major results of this study indicate that interests and a counterplan to reduce patient dose during mammography should be considered. Based on this study, the authors of this study will continue to measure exposure dose to set a new standard for patient exposure dose during mammography.

유방 X선촬영검사에서 환자가 받는 피폭선량과 장치의 이용 현황을 조사하기 위하여, 대한병원협회에 등록된 전국의 278개 의료기관(종합병원)에 설문지를 발송하여 설문에 응답한 161개 의료기관의 자료를 분석하였다. 조사 결과, 평균유선선량은 0.5 mGy 이하 14.9%, $0.51{\sim}1.0\;mGy$ 8.6%, $1.01{\sim}1.5\;mGy$ 14.9%, $1.51{\sim}2.0\;mGy$ 11.1%, $2.01{\sim}2.5\;mGy$ 9.8%, $2.51{\sim}3.0\;mGy$ 33.3%, 3.01 mGy 이상 7.4%이었다. 92.6%가 3 mGy 이하로 조사되었으며, 식품의약품안정청이 정한 평균유선선량 한도치인 3 mGy 근사값을 응답한 곳이 많았다. 최근, ICRP 등 방사선 방어 관련 국제기구는 유방 X선 촬영에서 평균유선선량을 성인 환자(50%지방, 50%유선)에 대해서 Mo target+Mo filter, 필름/스크린 시스템, 상하방향 촬영에서 4.2 cm로 압박된 유방에 대해서 1방향 3 mGy 이하로 제시하고 있으며, 점차 2 mGy, 또는 1.5 mGy로 낮추어 엄격하게 권고하고 있다. 따라서, 우리나라에서 유방 X선 촬영검사를 시행함에 있어 환자에 대한 피폭선량을 감소하기 위한 관심과 대책이 절실하게 요구된다. 아울러, 이번 조사를 바탕으로 의료기관의 유방 X선촬영 시 피폭선량을 실측하여 우리나라의 유방 X선촬영 시 환자피폭선량의 기준을 개발하기 위한 연구를 계속할 예정이다.