Three-dimensional Stress Analysis of Implant Systems with Micro Threads in the Maxillary Bone

다양한 마이크로쓰레드(Micro thread)의 개수를 가지는 임플란트의 상부구조물 형상과 하중조건에 따른 3차원 유한요소해석을 이용한 하악골의 응력분포에 관한 연구

  • 신하식 (연세대학교 대학원 기계공학과) ;
  • 한종현 (연세대학교 치과대학 보철학교실) ;
  • 이수홍 (연세대학교 기계공학부) ;
  • 전흥재 (연세대학교 기계공학부)
  • Published : 2005.03.01

Abstract

A comparative study of stress distributions in the maxillary bone with three different types of abutment was conducted. Finite element analysis was adopted to determine stress generated in the bone with the different implant systems with micro threads (Onebody type implant, Internal type implant, and External type implant). It was found that the types of abutments and the number of micro threads have significant influence on the stress distribution in the maxillary bone. They were due to the difference in the load transfer mechanism and the size of contact area between abutment and fixture. Also the maximum effective stress in the maxillary bone was increased with increasing inclination angle of load. It was concluded that the maximum effective stress in the bone was the lowest by the internal implant among the maximum effective stresses by other two types of implants and by appropriate number of micro threads, and that the specific number of micro thread was existed to decrease the maximum effective stress in the maxillary bone due to different implant systems and loading conditions.

Keywords

References

  1. Misch, C. E., 'Contemporary Implant Dentistry,' ed. 2 Missouri, Mosby, 1999
  2. Holmes, D. C., William, R. G., Vijay, K. G., John, C. K., 'Comparison of Stress Transmission in the IMZ Implant System with Polyoxymethylene or Titanium Intranmobile Element : A Finite Element Stress Analysis,' Int. J. Oral Maxillofac Implants, Vol. 7, pp. 450-458, 1992
  3. Rieger, M.R., Adams, W. K., Kinzel, G. L. and Brose, M.D., 'Alternative Materials for Three Endosseous Implants,' J. Prosth. Dent., p. 317,1989
  4. Rieger, M. R., Adams, W. K., Kinzel, G. L., 'A Finite Element Survey of Sleven Endosseous Implants,' J. Prosth. Dent., Vol. 63, pp. 457-465, 1990 https://doi.org/10.1016/0022-3913(90)90238-8
  5. Clelland, N. L., Ismail, Y. H., Zaki, H. S., Pipko, D., 'Three-dimensional Finite Element Stress Analysis in and around the Screw-Vent Implant,' Int. J. Oral Maxillofac. Implants, Vol. 6, pp. 391-398, 1991
  6. Siegele, D., 'Numerical Investigations of the Influence of Implant Shape on Stress Distributon in the Jaw Bone,' Int. J. Oral Maxillofac. Implants, Vol. 4,pp. 333-340,1989
  7. Pierrisnard, L., Hure, G., Barquins, M, Chappard, D., 'Two Dental Implants Designed for Immediate Loading: A Finite Element Analysis,' Int. J. Oral Maxillofac. Implants, Vol. 17, pp. 353 - 362, 2002
  8. Chun, H. -J, Cheong, S. -Y., Han, J.-H., Heo, S.-J., Chung, J.-P., Rhyu, I.-C., Choi, Y -C., Paik, H.-K., Ku, Y., Kim, M., 'Evaluation of Design Parameters of Osseointegrated Dental Implants Using Finite Element Analysis,' J. of Oral Rehabilitation, Vol. 29, pp.565-574,2002 https://doi.org/10.1046/j.1365-2842.2002.00891.x
  9. Holmgren, E. P., Seckinger, R. J., Kilgren, L. M., 'Evaluating Parameters of Osseo-intergated Dental Implant Using Finite Element Analysis a Two-dimensional Comparative Study Examining the Effects of Implant Diameter, Implant Shape, and Load Direction,' J. of Oral Implantology, Vol. 24, 1998
  10. Ziennkiewics, O. C., 'The Finite Element Method in Engineering Science,' ed. 4 New York, McGrawHill,1989
  11. Tanne, K., Burstone, C. J., Sakuda, M., 'Biomechanical Responses of Tooth Associated with Different Root Lengths and Alveolar Bone Height: Changes of Stress Distributions in the PDL,' J. Osaka Univ. DentSchool 29:17-24, 1989
  12. Boyer, R., Welsch, G. and Collings, E. W., 'Materials Properties Handbook: Titanium Alloys,' eds. ASM International, Materials Park, OH, 1994
  13. Merz, B. R., Hunenbart, S., 'Mechanics of the Implant-Abutment Connection: An 8-Degree Taper Compared to a Butt Joint Connection,' Int. J. Oral Maxillofac. Implants, Vol. 15, pp. 519-526, 2000