Assessment of Blood Glucose Values Measured by Blood Gas Analyzer or Portable Glucometer

혈액가스 분석기와 휴대용 혈당측정기로 측정된 혈당 수치의 평가

Park, Hue-Jung;Park, Chul-Soo;Park, Chong-Min;Ryu, Keon-Hee;Chang, Hae-Wone;Cho, Eun-Jeong;Lee, Yoon-Ki
박휴정;박철수;박종민;유건희;장혜원;조은정;이윤기

  • Published : 20060500

Abstract

Background: A portable glucometer is commonly used to immediately check the blood glucose level. In the anesthetic field, some blood gas analyzers can also give a rapid indication of the blood sugar level but the accuracy is unknown. Therefore, this study assessed the accuracy of the blood glucose values measured by either a blood gas analyzer or portable glucometer. Methods: Venous blood from diabetic patients was used to measure the glucose level with either a blood gas analyzer or a portable glucometer. The difference and 5% deviation from reference values was analyzed. These values were also assessed using a Bland-Altman plot and clinical significance was examined using a Clarke error grid. Results: The differences from the reference values were smaller using the blood gas analyzer (1.3 $\pm$ 7.8 mg/dl) than using the portable glucometer (-5.1 $\pm$ 16.7 mg/dl)(P < 0.01). 73.4% of the values measured by the blood gas analyzer and 40.0% of those measured by the portable glucometer were within 5% of the reference value. The 95% limits of agreement in the difference ranged from -14.3 to 16.9 in the blood gas analyzer and -38.5 to 28.2 in the portable glucometer. Error grid analysis showed that 100% of the values measured by the blood gas analyzer were located in zone A. When locating the values measured using the portable glucometer, 95.6% were located in zone A, and the remaining 4.4% are located in zone B. Conclusions: The blood gas analyzer measures the blood glucose more accurately than the portable glucometer. However, the blood glucose values measured by the portable glucometer are clinically acceptable.

Keywords

References

  1. Sacks DB, Bruns DE, Goldstein DE, Maclaren NK, McDonald JM, Parrott M: Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Clin Chem 2002; 48: 436-72
  2. American Diabetes Association: Self-monitoring of blood glucose. Diabetes Care 1996; 19: S62-6
  3. Antcliff RJ, Hugkulstone CE: The effect of fluorescein on the accuracy of blood glucose determination by glucose meters. Acta Ophthalmol Scand 1999; 77: 197-9 https://doi.org/10.1034/j.1600-0420.1999.770216.x
  4. Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 327: 307-10 https://doi.org/10.1016/S0140-6736(86)90837-8
  5. Clarke WL, Cox D, Gonder-Frederick LA, Carter W, Pohl SL: Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care 1987; 10: 622-8 https://doi.org/10.2337/diacare.10.5.622
  6. Mazzaferri EL, Skillman TG, Lanese RR, Keller MP: Use of test strips with colour meter to measure blood-glucose. Lancet 1970; 14: 331-3.
  7. Genter PM, Ipp E: Accuracy of plasma glucose measurements in the hypoglycemic range. Diabetes Care 1994; 17: 595-8 https://doi.org/10.2337/diacare.17.6.595
  8. Albertson C, Davis C, Ellison J, Chu C: Clinical evaluation of a new, miniaturized biosensor for self-monitoring of blood glucose. Clin Chem 1998; 44: 2056-7
  9. Tang Z, Louie RF, Lee JH, Lee DM, Miller EE, Kost GJ: Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing. Crit Care Med 2001; 29: 1062-70 https://doi.org/10.1097/00003246-200105000-00038
  10. Passey RB, Gillum RL, Fuller JB, Urry FM, Giles ML: Evaluation and comparison of 10 glucose methods and the reference method recommended in the proposed product class standard (1974). Clin Chem 1977; 23: 131-9
  11. Pelletier O, Arratoon C: Precision of glucose measurements in control sera by isotope dilution/mass spectrometry: proposed definitive method compared with a reference method. Clin Chem 1987; 33: 1397-402
  12. Chen ET, Nichols JH, Duh SH, Hortin G: Performance evaluation of blood glucose monitoring devices. Diabetes Technol Ther 2003; 5: 749-68 https://doi.org/10.1089/152091503322526969
  13. Kurahashi K, Maruta H, Usuda Y, Ohtsuka M: Influence of blood sample oxygen tension on blood glucose concentration measured using an enzyme-electrode method. Crit Care Med 1997; 25: 231-5 https://doi.org/10.1097/00003246-199702000-00006
  14. Cox DJ, Gonder-Frederick LA, Kovatchev BP, Julian DM, Clarke WL: Understanding error grid analysis. Diabetes Care 1997; 20: 911-2 https://doi.org/10.2337/diacare.20.6.911