Performance Characteristics of High Efficiency Fluid and Blood Warmer using Print Circuit Board Heater at Various Flow Rates

전자회로기판 히터를 이용한 고열효율 휴대용 수액 혈액 가온기의 유속에 따른 가온능력 및 안정성 평가

Jung, Sung-Won;Han, Tae-Hyung;Lee, Jin-Young;Kwak, In-Suk;Jung, Mi-Hwa;Won, Rim-Soo;Choi, Young-Ryong
정성원;한태형;이진영;곽인숙;정미화;원임수;최영룡

  • Published : 20061100

Abstract

Background: Currently available warming devices are often heavy and cumbersome, requiring development of more portable, user friendly, high efficiency fluid and blood warmer. The intravenous fluid heating capabilities of a new, heat-plate print circuit board (PCB) based warmer at various low flow rates were tested. Methods: Model Joyother BM-1$^{\circledR}$ was investigated for normal saline (0.9% NaCl) and colloid at various infusion rates (60-6000 ml/hr). Final temperatures were measured by electronic thermometer after passing through the warmer. Effective warming was defined as a fluid temperature $\geq$ 32$^{\circ}C$. Ambient temperature was maintained at 22-25$^{\circ}C$. Degree of heating capability and temperature decrease were compared and correlated between different flow rates. Results: The device warmed the room temperature crystalloid and colloid efficiently. Its warming capability was continuously improved as the flow rates increased in all tested flow rates, reaching maximum 41$^{\circ}C$. After the warming, the temperature decrease showed high negative correlation with the flow rates. The extent of cooling was less in colloid. No overheating was noted at sudden brake. Conclusions: Joyother BM-1$^{\circledR}$ heated crystalloid and colloid sufficiently and safely enough for clinical application (exit temperature $\geq$ 35$^{\circ}C$) at various flow rates. The warming capacity and the length of the intravenous tube infusion system determined the efficiency of the warmer. Further study is warranted for the efficiency of warming for the blood and its element changes.

Keywords

References

  1. Aslam AF, Aslam AK, Vasavada BC, Khan IA: Hypothermia: evaluation, electrocardiographic manifestations, and management. Am J Med 2006; 119: 297-301 https://doi.org/10.1016/j.amjmed.2005.09.062
  2. Arthurs Z, Cuadrado D, Beekley A, Grathwohl K, Perkins J, Rush R, et al: The impact of hypothermia on trauma care at the 31st combat support hospital. Am J Surg 2006; 191: 610-4 https://doi.org/10.1016/j.amjsurg.2006.02.010
  3. Martin RS, Kilgo PD, Miller PR, Hoth JJ, Meredith JW, Chang MC: Injury-associated hypothermia: an analysis of the 2004 National Trauma Data Bank. Shock 2005; 24: 114-8 https://doi.org/10.1097/01.shk.0000169726.25189.b1
  4. Brieva J, McFadyen B, Rowley M: Severe hypothermia: challenging normal physiology. Anaesth Intensive Care 2005; 33: 662-4
  5. Fildes J, Fisher S, Sheaff CM, Barrett JA: Effects of short heat exposure on human red and white blood cells. J Trauma 1998; 45: 479-84 https://doi.org/10.1097/00005373-199809000-00008
  6. Rickard CM, Couchman BA, Hughes M, McGrail MR: Preventing hypothermia during continuous veno-venous haemodiafiltration: a randomized controlled trial. J Adv Nurs 2004; 47: 393-400 https://doi.org/10.1111/j.1365-2648.2004.03117.x
  7. Ohtsuka N, Yamakage M, Chen X, Kamada Y, Namiki A: Evaluation of four techniques of warming intravenous fluids. J Anesth 2002; 16: 145-9 https://doi.org/10.1007/s005400200010
  8. Moerer O, Brauer A, Weyland W, Braun U: Warming efficacy and blood damaging of blood and infusion warmers. Anasthesiol Intensivmed Notfallmed Schmerzther 2004; 39: 138-46 https://doi.org/10.1055/s-2004-814326
  9. Schnoor J, Weber I, Macko S, Heussen N, Rossaint R: Heating capabilities of the Hotline and Autoline at low flow rates. Paediatr Anaesth 2006; 16: 410-6 https://doi.org/10.1111/j.1460-9592.2005.01811.x
  10. Hirsch J, Menzebach A, Welters ID, Dietrich GV, Katz N, Hempelmann G: Indicators of erythrocyte damage after microwave warming of packed red blood cells. Clin Chem 2003; 49: 792-9 https://doi.org/10.1373/49.5.792
  11. Russell WJ: A review of blood warmers for massive transfusion. Anaesth Intensive Care 1974; 2: 109-30
  12. Dubick MA, Brooks DE, Macaitis JM, Bice TG, Moreau AR, Holcomb JB: Evaluation of commercially available fluid-warming devices for use in forward surgical and combat areas. Mil Med 2005; 170: 76-82 https://doi.org/10.7205/MILMED.170.1.76
  13. Presson RG Jr, Bezruczko AP, Hillier SC, McNiece WL: Evaluation of a new fluid warmer effective at low to moderate flow rates. Anesthesiology 1993; 78: 974-80 https://doi.org/10.1097/00000542-199305000-00023
  14. Bissonnette B, Paut O: Active warming of saline or blood is ineffective when standard infusion tubing is used: an experimental study. Can J Anaesth 2002; 49: 270-5 https://doi.org/10.1007/BF03020526
  15. Faries G, Johnston C, Pruitt KM, Plouff RT: Temperature relationship to distance and flow rate of warmed i.v. fluids. Ann Emerg Med 1991; 20: 1198-200 https://doi.org/10.1016/S0196-0644(05)81470-2
  16. Smallman JM, Morgan M: Evaluation of the Level 1 Hotline blood warmer. Anaesthesia 1992; 47: 869-71 https://doi.org/10.1111/j.1365-2044.1992.tb03152.x
  17. Eastlund T, Van Duren A, Clay ME: Effect of heat on stored red cells during non-flow conditions in a blood-warming device. Vox Sang 1999; 76: 216-9 https://doi.org/10.1046/j.1423-0410.1999.7640216.x
  18. Lee JH, Mintz PD: A method for estimating the delivery temperature of intravenous fluids. Anesth Analg 1994; 79: 155-9
  19. Hill HR, Oliver CK, Lippert LE, Greenwalt TJ, Hess JR: The effects of polyvinyl chloride and polyolefin blood bags on red blood cells stored in a new additive solution. Vox Sang 2001; 81: 161-6 https://doi.org/10.1046/j.1423-0410.2001.00097.x
  20. Muylle L, Vanderplanken M, Goossens W, Stewart M, Payrat JM: Storage of saline-adenine-glucose-mannitol suspended red cells in diethylhexyl phtalate and butyryl-n-trihexyl-citrate plasticized polyvinyl chloride containers. An in vitro comparative study. Transfus Sci 1994; 15: 163-9 https://doi.org/10.1016/0955-3886(94)90091-4
  21. Seidl S, Gosda W, Reppucci AJ: The in vitro and in vivo evaluation of whole blood and red cell concentrates drawn on CPDA-1 and stored in a non-DEHP plasticized PVC container. Vox Sang 1991; 61: 8-13 https://doi.org/10.1111/j.1423-0410.1991.tb00919.x