The reliability of the cephalogram generated from cone-beam CT

Cone-beam CT로부터 제작된 측모 두부계측방사선사진의 정확도 평가

  • Kang, Ji-Young (Department of Orthodontics, School of Dentistry, Chosun University) ;
  • Kim, Kwang-Won (Department of Orthodontics, School of Dentistry, Chosun University) ;
  • Lim, Sung-Hoon (Department of Orthodontics, School of Dentistry, Chosun University)
  • 강지영 (조선대학교 치과대학 교정학교실) ;
  • 김광원 (조선대학교 치과대학 교정학교실) ;
  • 임성훈 (조선대학교 치과대학 교정학교실)
  • Published : 2007.12.31

Abstract

Three-dimensional approaches for the diagnosis and analysis of the dentofacial area are becoming more popular in accordance with the development of cone-beam CT (CBCT). The purposes of this study were to evaluate the reliability of cephalometric measurements of lateral cephalograms generated from a CBCT image by making comparisons with the traditional digital lateral cephalogram, and to evaluate the possibility of the clinical application of CBCT generated cephalogram images. Methods: Twenty patients whose external auditory meatus could be identified in the CBCT image were selected, and both CBCT and digital cephalograms were taken. Differences between the measurements of both cephalograms were tested by paired t-test. Results: Among the 22 measurements used, only U1-FH, Mx6 to PTV, and maxillomandibular difference showed statistically significant differences between the CBCT generated cephalogram and the digital cephalogram. Conclusions: The results suggest that the CBCT generated cephalogram can be used for some cephalometric measurements not requiring porion, PTV, condylion as a landmark (SNA, SNB, U1 to SN, IMPA, interincisal angle, etc.).

최근 computed tomography (CT) 데이터로부터 실제 두부계측방사선사진 촬영기에서 촬영한 사진과 동일한 확대율을 갖는 두부계측방사선사진을 형성할 수 있는 기능을 갖춘 소프트웨어(Accurex, CyberMed, Seoul, Korea)가 소개되었다. 이에 본 연구에서는 cone-beam CT (이하 CBCT) 데이터로부터 이 소프트웨어를 이용하여 제작된 측모 두부계측방사선사진과 기존의 두부계측방사선사진촬영기로 촬영된 측모 두부계측방사선사진에서 각각 두부계측방사선사진 분석을 시행하여 그 차이를 비교함으로써 소프트웨어를 통해 CBCT로부터 제작된 측모 두부계측방사선사진의 활용 가능성을 평가하고자 하였다. 20명의 교정 환자를 대상으로 CBCT와 측모 두부계측방사선사진을 동시에 촬영하였다. CBCT로부터 제작된 측모 두부계측방사선사진과 기존 방식의 측모 두부계측방사선사진에서 두부계측 방사선사진 분석을 시행한 후 두 사진의 계측치의 차이를 paired t-test로 검정하였다. 23개의 계측치들 중, 20개 항목에서는 통계학적으로 유의한 차이가 관찰되지 않았으나, U1-FH, Mx6 to PTV, Maxillomandibular difference의 항목에서는 유의한 차이가 나타났다 이러한 차이는 CBCT로부터 형성된 측모 두부계측방사선사진의 해상도가 낮아 일부 해부학적 지표를 확인하기 어렵기 때문이며, 임상에서 CBCT로부터 제작된 측모 두부계측방사선사진을 porion, PTV, condylion과 같은 일부 해부학적 지표를 이용하지 않는 SNA, SNB, U1 to SN, IMPA, interincisal angle 등 흔히 사용되는 간단한 측모 두부계측방사선사진 계측에 활용 가능할 것이다.선 병용요법은 절제 불가능한 원발성 간암의 증상호전 및 국소적 치료에 효과가 있는 것으로 생각되며 생존율 및 부작용 등에 관해서는 앞으로 연구가 더 진행되어야 할 것으로 생각된다.용 선형가속기의 전자선 방출구에 장착된 산란판과 조사면을 조정하는 cone에 의하여 발생되는 저 에너지 산란전자선이 alanine dosimeter에 측정된 것으로서 에너지가 증가될수록 오염 정도가 증가되었다. 본 실험을 통하여 지금까지 고에너지 전자선량계측에서 전리상에 의한 전기량 측정과 산란선이 없는 단일 에너지로만 간주하여 계산하였던 전자선 흡수선량 측정방법을 직접 흡수선량 측정이 가능한 Alanine/ESR dosimetry로서 교정하는 것이 바람직하다고 생각한다.$6\%$, 수술이 필요하였던 환자는 $1\%$로 합병증의 증가 없이 A점에 8900cGy, 직장에 7300cGy정도의 국소치료가 가능함을 보여 주었으나 생존율에 대한 추후 분석이 따라야 할 것으로 사료된다.보면 6,000rad이하에서 20/63$(34.9\%)$의 실패율에 비하여 6,000rad이상일 때는 10/49$(20.4\%)$의 실패율이었다. 연령 벨로는 $40\~49$세에서 실패율(14/41 $24.1\%$)이 많았다. 본 성적으로 보아서 생존율은 여러 저자들과 큰 차이가 없음을 알 수 있었고 A점 선량은 8,000rad 이상, B점은 6,000rad이상이 조사되어야 적정선량이 됨을 시사해 주고 있다.\%$가 대부분이 하루 세끼 식사를 규칙적으로 하고 있었으며 식사속도는 허겁지겁 빨리 섭취하는 경우가 남자는 $31.0\%$,

Keywords

References

  1. Broadbent BH. A new x-ray technique and its application to orthodontia. Angle Orthod 1931;1:45-66
  2. Wylie WH, Elasser WA. Undistorted vertical projections of the head from lateral and postero-anterior roentogenograms. Am J Roentgenol 1948;60:414-17
  3. Vogel CJ, Correction of frontal dimensions from head x-rays. Angle Orthod 1967;37:1-8
  4. Grayson BH, McCarthy JG, Bookstein F. Analysis of craniofacial asymmetry by multiplane cephalometry. Am J Orthod 1983;84:217-24 https://doi.org/10.1016/0002-9416(83)90129-X
  5. Grayson BH, Cutting C, Bookstein FL, Kim H, McCarthy JG. The three-dimensional cephalogram: theory, technique, and clinical application. Am J Orthod Dentofacial Orthop 1988;94:327-37 https://doi.org/10.1016/0889-5406(88)90058-3
  6. Bookstein FL, Grayson BH, Cutting C, Kim H, McCarthy JG. Landmarks in three-dimensions: reconstruction from cephalogram versus direct observation. Am J Orthod Dentofacial Orthop 1991;100:133-40 https://doi.org/10.1016/S0889-5406(05)81520-3
  7. Brown T, Abbott AH. Computer-assisted location of reference points in three dimensions for radiographic cephalometry. Am J Orthod Dentofacial Orthop 1989;95:490-8 https://doi.org/10.1016/0889-5406(89)90412-5
  8. Baumrind S, Moffitt FH, Curry S. Three-dimension x-ray stereometry from paired coplanar images. A progress report. Am J Orthod 1983;84:292-312 https://doi.org/10.1016/S0002-9416(83)90346-9
  9. Chang HS, Baik HS. A proposal of landmarks for craniofacial analysis using three-dimensional CT imaging. Kor J Orthod 2002;32:313-25
  10. Swennen GRJ, Barth EL, Schutyser F, De Groeve P, Lemaitre A. Three-dimensional (3-D) cephalometry, the basics for virtual planning. J Cranio Maxillofac Surg 2004;32:(Suppl 1)135 https://doi.org/10.1016/j.jcms.2004.01.005
  11. Swennen GRJ. Three-dimensional cephalometric reference system. In: Swennen GRJ, Schutyser F and Hausamen JE, Editors, Three-dimensional cephalometry. A color atlas and manual. Heidelberg: Spinger 2005;94-112
  12. Swennen GR, Schutyser F. Three-dimensional cephalometry: Spiral multi-slice vs cone-beam computed tomography. Am J Orthod Dentofacial Orthop 2006;130:410-6 https://doi.org/10.1016/j.ajodo.2005.11.035
  13. Feldkamp LA, Davis LC, Kress JW. Practical conebeam algorithm. J Opt Soc Am 1984;1:612-9 https://doi.org/10.1364/JOSAA.1.000612
  14. Mah J, Danforth RA, Bumann A, Hatcher D. Radiation absorbed in maxillofacial imaging with a new dental computed tomography device. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2003;96:508-13 https://doi.org/10.1016/S1079-2104(03)00350-0
  15. Tsiklakis K, Donta C, Gavala S, Karayianni K, Kamenopoulou V, Hourdakis CJ. Dose reduction in maxillofacial imaging using low dose cone beam CT. Eur J Radiol 2005;56:413-7 https://doi.org/10.1016/j.ejrad.2005.05.011
  16. Halazonetis DJ. From 2-dimensional cephalograms to 3-dimensional computed tomography scans. Am J Orthod Dentofacial Orthop 2005;127:627-37 https://doi.org/10.1016/j.ajodo.2005.01.004
  17. Huang J, Bumann A, Mah J. Three-dimensional radiographic analysis in orthodontics. J Clin Orthod 2005;39:421-8
  18. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;327:307-10 https://doi.org/10.1016/S0140-6736(86)90837-8
  19. Demura N, Tsurusako Y, Segami N. Characteristics of digital cephalograms and film/screen cephalograms: A comparative study. World J Orthod 2001;2:350-5
  20. Schulze RK, Gloede MB, Doll GM. Landmark identification on direct digital versus film-based cephalometric radiographs: A human skull study. Am J Orthod Dentofacial Orthop 2002;122:635-42 https://doi.org/10.1067/mod.2002.129191
  21. Ross LL, Munn MR. Comparing digital serial cephalogram images for growth or treatment changes. Am J Orthod Dentofacial Orthop 2005;128:161-2 https://doi.org/10.1016/j.ajodo.2005.02.013
  22. Farman AG, Scarfe WC. Development of imaging selection criteria and procedures should precede cephalometric assessment with cone-beam computed tomography. Am J Orthod Dentofacial Orthop 2006;130:257-65 https://doi.org/10.1016/j.ajodo.2005.10.021
  23. Moshiri M, Scarfe WC, Hilgers ML, Scheetz JP, Silveira AM, Farman AG. Accuracy of linear measurements from imaging plate and lateral cephalometric images derived from cone-beam computed tomography. Am J Orthod Dentofacial Orthop 2007;132:550-60 https://doi.org/10.1016/j.ajodo.2006.09.046
  24. Mah J, Hatcher D. Three-dimensional craniofacial imaging. Am J Orthod Dentofacial Orthop 2004;126:308-9 https://doi.org/10.1016/j.ajodo.2004.06.024