Objective Measurement of Accommodative Responses with Open-Field Autorefractor

개방형 자동굴절계를 이용한 조절반응의 타각적 측정

Han, Gyeong-Ae;Hwang, Jeong-Hee;Mah, Ki-Choong
한경애;황정희;마기중

  • Published : 2009.03.31

Abstract

Purpose : This study measured objectively the accommodative responses stimulated by push-up method and minus lens method with the NVision-K 5001 open-field autorefractor, and compared the results with previously published data. Methods : Forty-five subjects aged 19 to 29 years (mean $22.7{\pm}2.4$ years) participated in this study. Three conditions were used to stimulate accommodation and accommodative responses were measured objectively by NVision-K 5001 : (1) push-up test under binocular viewing condition ; the target was placed at vergence ranging from 0-5D in 1D steps. (2) push-up test under monocular viewing condition ; the target was placed at vergence ranging from 1-5D in 1D steps. (3) minus lens test under monocular viewing condition ; a distant(6 m) target viewed through increasing powered minus lenses from 1-5D in 1D steps. Trial lens calibration procedures were also performed to assess the accuracy of the NVision-K 5001. Results : The calibration of NVision-K 5001 with trial lenses showed linear regression equation which was close to the ideal 1 : 1 line(y=-1.005x + 0.060, $r^2$=0.999, p<0.001). The accommodative responses measured with the NVision-K 5001 for the push-up and minus lens-stimulated accommodation progressively lagged behind the increasing accommodative stimulus. Both push-up methods under monocular and binocular viewing conditions showed significantly higher accommodative responses than minus lens method(p<0.001). Push-up method under binocular viewing condition demonstrated slightly higher accommodative responses compared with monocular condition, but the difference was not significant(p = 0.958). Conclusion : The objective accommodative responses stimulated by push-up and minus lens method and measured with open-field autorefractor showed good agreement with previously published data.

목 적 : 개방형 자동굴절계인 NVision-K 5001을 이용하여 Push-up법과 Minus lens법으로 자극된 조절반응을 타각적으로 측정하여 기존 연구와 비교하였다. 방 법 : 본 연구의 취지에 동의하는 19~29세 사이(평균 $22.7{\pm}2.4$세)의 45명을 대상자로 선정하였다. 세 가지 조건으로 조절을 자극하고 NVision-K 5001을 이용하여 조절반응량을 타각적으로 측정하였다 : (1) 양안시 상태에서 push-up 검사 ; 타깃을 조절자극 0~5D 범위에서 1D 간격으로 위치시켰다. (2) 단안시 상태에서 push-up 검사 ; 타깃을 조절자극 0~5D 범위에서 1D 간격으로 위치시켰다. (3) 단안시 상태에서 minus lens 검사 ; 원거리(6 m) 타깃을 주시시키고 마이너스 시험렌즈를 1~5D까지 1D 간격으로 증가시켰다. NVision-K 5001의 정확도를 검사하기 위해 시험렌즈를 이용한 calibration을 수행하였다. 결 과 : 시험렌즈를 이용하여 수행한 Nvision-K 5001의 calibration 결과는 이상적인 1 : 1선에 근접한 선형회귀식을 보였다(y=-1.005x + 0.060, $r^2$=0.999, p< 0.001). Pushup법과 minus lens법으로 조절을 자극하고 NVision-K 5001을 이용하여 조절반응량을 측정한 결과, 조절자극의 증가량에 비하여 조절반응의 증가량이 점차 감소하는 유형을 보였다. 양안시 및 단안시 상태에서 push-up법으로 측정한 조절반응은 minus lens법으로 측정한 값보다 유의하게 높았다(p<0.001). 양안시 상태의 push-up법으로 측정한 조절반응은 단안시 상태의 push-up법으로 측정한 값보다 약간 높았지만 유의성은 없었다(p = 0.958). 결 론 : Push-up법과 minus lens법으로 조절을 자극하고 개방형 자동굴절계로 측정한 조절반응량은 선행 연구결과와 유사하였다.

Keywords