Clinical Outcome of Diffractive Multifocal Aspheric Intraocular Lens

회절 방식의 다초점 비구면 인공수정체의 단기 임상 결과

Chang, Min-Wook;Eom, Young-Sub;Kang, Su-Yeon;Kim, Kyun-Hyung;Song, Jong-Suk;Kim, Hyo-Myung
장민욱;엄영섭;강수연;김균형;송종석;김효명

  • Published : 2009.04.15

Abstract

Purpose: To compare the clinical outcome between Tecnis ZM900 and Tecnis ZA9003. Methods: We reviewed 20 eyes of 11 patients implanted with Tecnis ZM900 and 20 eyes of 20 patients implanted with Tecnis ZA9003. The clinical outcomes of these two intraocular lenses were evaluated one month after operation and consisted of distant, intermediate, and near visual acuity; depth of focus; contrast sensitivity; wavefront aberration; and patient satisfaction. Six months after the operation a reevaluation was performed for the group who had received the Tecnis ZM 900 implant. Results: Near and intermediate vision was better in the Tecnis ZM900 group. There were no statistical differences between the two groups with respect to distant vision and wavefront aberration. Contrast sensitivity was better in the Tecnis ZA9003 group and depth of focus was deeper in the Tecnis ZM900 group. There were no statically significant differences in the result between one month and six month after the operation. Conclusions: The Tecnis multifocal ZM900 IOL can be effective at improving patient satisfaction after cataract surgery as well as for correcting presbyopia.

목적: 회절 방식의 다초점 비구면 인공수정체 Tecnis ZM900과 단초점 비구면 인공수정체 Tecnis ZA9003 삽입 후 수술 결과를 비교하고자 하였다. 대상과 방법: 다초점군 11명 20안, 단초점군 20명 20안을 대상으로 수술 1개월 후 시력, 대비감도, 고위수차, 초점심도를 비교하였고, 다초점군에서는 양안과 단안의 시력, 대비감도 및 초점심도를 비교하였으며, 환자 만족도를 조사하였고, 수술 6개월간 추적 관찰하여 수술 1개월 후 결과와 비교하였다. 결과: 다초점군에서 근거리 및 중간거리 시력이 더 좋았고(p<0.05) 원거리 시력은 유의한 차이가 없었다. 대비감도는 단초점군이 더 좋았고(p<0.05), 초점심도는 다초점군에서 더 깊었으며(p<0.05) 고위수차는 두 군간의 유의한 차이는 없었다. 다초점군에서 단안 및 양안의 시력, 대비감도, 초점심도를 비교한 결과 통계적으로 유의한 차이는 없었다. 다초점군의 6개월 관찰 결과 1개월 후 결과와 유의한 차이는 없었다. 결론: Tecnis 다초점 인공수정체는 백내장 수술 환자만족도 개선에 도움이 될 것으로 생각하며 노안 교정에도 유용할 것으로 생각된다.

Keywords

References

  1. Martinez Palmer A, Gomez Faina P, Espana Albelda A, et al. Visual function with bilateral implantation of monofocal and multifocal intraocular lenses. J Refract Surg 2008;24:257-64
  2. Goes FJ. Refractive lens exchange with the diffractive multifocal tecnis ZM900 intraocular lens. J Refract Surg 2008;24:243-50
  3. Fernandez-Vega L, Alfonso JF, Rodriguez PP, Montes-Mico R. Clear lens extraction with multifocal apodized intraocular lens implantation. Ophthalmology 2007;114:1491-8 https://doi.org/10.1016/j.ophtha.2006.10.058
  4. Fernandez-Vega L, Alfonso JF, Montes-Mico R, Amhaz H. Visual acuity tolerance to residual refractive errors in patients with an apodized diffractive intraocular lens. J Cataract Refract Surg 2008;34:199-204 https://doi.org/10.1016/j.jcrs.2007.10.020
  5. Lee DY, Roh JH, Shyn KH. Current trends in cataract surgery in Korea – 2005 survey for KSCRS members. J Korean Ophthalmol Soc 2007;48:485-92
  6. Lane SS, Morris M, Nordan L, et al. Multifocal intraocular lenses. Ophthalmol Clin North Am 2006;19:89-105
  7. Hutz WW, Eckhardt HB, Rohrig B, Grolmus R. Reading ability with 3 multifocal intraocular lens models. J Cataract Refract Surg 2006;32:2015-21 https://doi.org/10.1016/j.jcrs.2006.08.029
  8. Kohnen T, Allen D, Boureau C, et al. European multicenter study of the AcrySof ReSTOR apodized diffractive intraocular lens. Ophthalmology 2006;113:584
  9. Blaylock JF, Si Z, Vickers C. Visual and refractive status at different focal distances after implantation of the ReSTOR multifocal intraocular lens. J Cataract Refract Surg 2006;32:1464-73 https://doi.org/10.1016/j.jcrs.2006.04.011
  10. Chiam PJ, Chan JH, Aggarwal RK, Kasaby S. ReSTOR intraocular lens implantation in cataract surgery: quality of vision. J Cataract Refract Surg 2006;32:1459-63 https://doi.org/10.1016/j.jcrs.2006.04.015
  11. Sallet G. Refractive outcome after bilateral implantation of an apodized diffractive intraocular lens. Bull Soc Belge Ophtalmol 2006;299:67-73
  12. Souza CE, Muccioli C, Soriano ES, et al. Visual performance of AcrySof ReSTOR apodized diffractive IOL: a prospective comparative trial. Am J Ophthalmol 2006;141:827-32 https://doi.org/10.1016/j.ajo.2005.12.031
  13. Bi H, Cui Y, Ma X, et al. Early clinical evaluation of Acrysof ReSTOR multifical intraocular lens for treatment of cataract. Ophthalmologica 2008;222:11-6 https://doi.org/10.1159/000109272
  14. Pepose JS, Qazi MA, Davies J, et al. Visual performance of patients with bilateral vs combination Crystalens, Rezoom, and ReSTOR intraocular lense implants. Am J Ophthalmol 2007;144:347-57 https://doi.org/10.1016/j.ajo.2007.05.036
  15. Vingolo EM, Grenga P, Iacobelli L, Grenga R. Visual acuity and contrast sensitivity: acrysof ReSTOR apodized diffractive vursus Acrysof SA60AT monofocal intraocular lenses. J Cataract Refract Surg 2007;33:1244-7 https://doi.org/10.1016/j.jcrs.2007.03.052
  16. Toto L, Falconio G, Vecchiarino L, et al. Visual performance and biocompatibility of 2 multifocal diffractive IOLs. J Catarct Refract Surg 2007;33:1419-25 https://doi.org/10.1016/j.jcrs.2007.05.010

Cited by

  1. One-year Outcome of Monocular Implant of Aspheric Multifocal IOL vol.51, pp.6, 2009, https://doi.org/10.3341/jkos.2010.51.6.822
  2. The Study on Target Refraction to Improve Visual Quality in Patients Implanted with Multifocal IOL vol.53, pp.2, 2009, https://doi.org/10.3341/jkos.2012.53.2.223
  3. Visual and Optical Functions after Diffractive Multifocal Intraocular Lens vol.53, pp.3, 2009, https://doi.org/10.3341/jkos.2012.53.3.396
  4. Clinical Results and Optical Quality of Diffractive Multifocal IOL Implantation after Myopic Refractive Surgery vol.55, pp.12, 2009, https://doi.org/10.3341/jkos.2014.55.12.1779
  5. Clinical Results and Optical Quality of Diffractive Multifocal Intraocular Lens vol.56, pp.12, 2015, https://doi.org/10.3341/jkos.2015.56.12.1867