Development of a Non-invasive Ultrasonic Measurement System for tissue elasticity

비침습적 초음파 조직 탄성도 측정 시스템 개발

  • Lee, G.J. (Department of Biomedical Engineering, Yonsei University) ;
  • Choi, W.H. (Department of Biomedical Engineering, Yonsei University) ;
  • Yu, J.W. (Department of Biomedical Engineering, Yonsei University) ;
  • Seo, J.B. (Department of Biomedical Engineering, Yonsei University) ;
  • Choi, S.H. (Hana traditional Oriental medicine hospital) ;
  • Shin, T.M. (Department of Biomedical Engineering, Yonsei University)
  • 이균정 (연세대학교 보건과학대학 의공학과) ;
  • 최우혁 (연세대학교 보건과학대학 의공학과) ;
  • 유재원 (연세대학교 보건과학대학 의공학과) ;
  • 서종범 (연세대학교 보건과학대학 의공학과) ;
  • 최서형 (연세대학교 보건과학대학 의공학과) ;
  • 신태민 (연세대학교 보건과학대학 의공학과)
  • Published : 2009.12.31

Abstract

Diseases caused by indurate tissues of the internal organs are liver cirrhosis and abdominal sclerosis. The cause of chronic gastro-intestinal disease is a digestive system disorder and a defecation disorder. They impede peristaltic movement and digestive system with the symptom that indurate tissues. The purpose of the present study was to determine the disease grade quantitatively by measuring an indurated standard of tissues and organs. For the measurement of elasticity, we designed the system that measure the displacement of the substance and approved pressure using ultrasound transducer. For verification of developed system, we compared elasticity as results of experiment between the developed system and public elasticity measurement machine at individual plastic phantoms made by plastic hardener and softener. Elasticity of the plastic phantoms is averagely 0.007MPa lower measured by developed system than Micro-indenter, and less than 10% errors. Comparing with economical value and accuracy between developed system and Micro-indenter, the system is significant of measurement for tissue elasticity. Thus, it is possible to measure a elasticity at human tissue and organ. A chronic gastro-disease as well as grade could be decided objective validity using this system.

Keywords

References

  1. Ki Tae Suk, Dae Wook Lim, Moon Young Kim, Dong Hun Park, Kyu Hong Kim, Jung Min Kim, Jae Woo Kim, Hyun Soo Kim, Sang Ok Kwon, Soon Koo Baik, Sung Jin Park, "Thickening of the Gastric Wall on Transabdominal Sonography: A Sign of Gastric Cancer", Journal of Clinical Ultrasound, Vol. 36 No. 8, pp 462-466, 2007
  2. Takaki M., "Gut pacemaker cells: The interstitial cells of Cajal (ICC)", J Smooth Muscle Res. J Smooth Muscle Res.,Vol. 39, No. 5, pp137-161, 2003 https://doi.org/10.1540/jsmr.39.137
  3. K. W. Min, and M. Leabu, "interstitial cells of cajal(ICC) and gastrointestinal stromal tumor(GIST): facts, speculations, and myths," J, Cell Mol Med.,vol. 10, pp.995-1013, 2006 https://doi.org/10.1111/j.1582-4934.2006.tb00541.x
  4. Ophir, J., CCspedcs, I., Ponnekanti, H., Yazdi, Y., et al.., "Elastography: a quantitative method for imaging the elasticity of biological tissues," Ultrasonic Imaging, Vol. 14, pp. I1 1-134, 1991
  5. Chen EJ, Novakofski J, Jenkins WK, O'Brien WD Jr. "Young's modulus measurement of soft tissues with application to elasticity imaging," IEEE trans. Ultrasonics, Ferroelectrics and Frequency control. 1996;43(1):191-194 https://doi.org/10.1109/58.484478
  6. M. O'Donnell, A. R.Skovoroda, B. M. Shapo, and S. Y. Emelianov," Internal displacement and strain imaging using ultrasonic speckle tracking," IEEE Trans. Ultrason., Ferroelect., Freq. Contr., vol. 41, pp. 314-325, May 1994 https://doi.org/10.1109/58.285465
  7. Mark A. Lubinski "Speckle Tracking Methods for Ultrasonic Elasticity Imaging Using Short-Time Correlation", IEEE Trans. Ultrason., Ferroelect., Freq. Contr. vol. 46, NO. 1, January 1999
  8. Zhang, M. Zhang, YP and Mak AFT. "Estimating the effective Young's modulus of soft tissues form indentation tests-nonlinear finite element analysis of effects of friction and large deformation" Med. Eng. Phys. Vol. 19, No.6 pp512-517, 1997 https://doi.org/10.1016/S1350-4533(97)00017-9
  9. P. Wellman, R. Howe, E. Dalton, and K. A. Kern, "Breast tissue stiffness in compression is correlated to histological diagnosis," Tech. Rep., Harvard BioRobotics Laboratory, Harvard University, Cambridge, Mass, USA, 1999
  10. 이균정, 박동희, 신태민, 서종범, "초음파 영상용 플라스틱 기반의 팬텀제작 및 특성분석" Journal of Biomedical Engineering Research, 29권, 4호 pp302-306, 2008