DOI QR코드

DOI QR Code

Mechanical safety evaluation of ceramic ball head for total hip replacement using finite element method

인공고관절 전치환술에서 세라믹 볼 헤드의 기계적 안정성 평가를 위한 유한요소 해석

  • Han, Sung-Min (Biomedical Research Center(Korea Institute of Science and Technology) ;
  • Chu, Jun-Uk (Biomedical Research Center(Korea Institute of Science and Technology) ;
  • Chun, Heoung-Jae (Mechanical Engineering Yonsei University) ;
  • Kim, Jung-Sung (Coren_tec Co. Ltd.) ;
  • Choi, Kui-Won (Biomedical Research Center(Korea Institute of Science and Technology) ;
  • Youn, In-Chan (Biomedical Research Center(Korea Institute of Science and Technology)
  • 한성민 (한국과학기술연구원 의과학연구센터) ;
  • 추준욱 (한국과학기술연구원 의과학연구센터) ;
  • 전흥재 (연세대학교 기계공학과) ;
  • 김정성 ((주)코렌텍) ;
  • 최귀원 (한국과학기술연구원 의과학연구센터) ;
  • 윤인찬 (한국과학기술연구원 의과학연구센터)
  • Received : 2010.05.10
  • Accepted : 2010.11.15
  • Published : 2010.12.31

Abstract

A ceramic articulating system in total hip replacement thought to be superior to metal-on-polyethylene due to its extremely low coefficient of friction and potential for high resistance to wear. But ceramic is brittle, which makes it mechanically and theoretically susceptible to fracture under certain mechanical conditions. In the current study, nine different models of ceramic ball heads were mechanically evaluated using 3D finite element(FE) analyses. It was found that the maximum stress in all ceramic models was lower than ceramic flexural strength, and it satisfied the requirements of the FDA Gaudience for artificial hip implant. Thus, ceramic ball head models introduced in the current study could be mechanically safe for clinical applications.

Keywords

References

  1. Y. S. Park, J. H. Kim, Y. H. Kim, and K. C. Park, "Osteonecrosis of Femoral Head after Pelvic Fracture," J Korean Orthop Assoc, vol. 44, pp. 495-498, 2009. https://doi.org/10.4055/jkoa.2009.44.4.495
  2. M. J. Fagan, and A. J. C. Lee, " Material selection in the design of the femoral component of cemented total hip replacements," Clinical Materials, vol. 1, no. 3, pp. 151-167, 1986. https://doi.org/10.1016/S0267-6605(86)80029-4
  3. B. Latham, and T. Goswami, "Effect of geometric parameters in the design of hip implants paper IV," Materials & Design, vol. 25, no. 8, pp. 715-722, 2004 . https://doi.org/10.1016/j.matdes.2004.01.012
  4. A. Ramos, and J. A. Simoes, "Tetrahedral versus hexahedral finite elements in numerical modelling of the proximal femur,", Medical Engineering & Physics, vol. 28, no. 9, pp. 916-924, 2005.
  5. B. Weisse, M. Zahner, W. Weber, and W. Rieger, "Improvement of the reliability of ceramic hip joint implants," Journal of Biomechanics, vol. 36, pp. 1633-1639, 2003. https://doi.org/10.1016/S0021-9290(03)00186-6
  6. V.C. Mow, and W. C. Hayes, "Basic Orthopaedic Biomechanics Second Edition," Lippincott-Raven, 1997.
  7. P. Helwig, C. Faust, U. Hindenlang, A. Hirschmuller, L. Konstantinidis, C. Bahrs, N. Sudkamp, and R. Schneider, "Finte element analysis of four different implants inserted in different positions to stabilize an idealized trochanteric femoral fracture," Injury, vol. 40, pp. 288-295, 2008.
  8. J. R. T. Jeffers, M. Browne, A. B. Lennon, P. J. Prendergast, and M. Taylor "Cement mantle fatigue in total hip replacement : Experimental and computational testing," Journal of Biomechanics, vol. 40, pp. 1525-1533, 2007. https://doi.org/10.1016/j.jbiomech.2006.07.029
  9. H. J. Kim, J. S. Lim, S. M. Han, J. H. You, K. W. Choi, and I. C. Yoon, "Evaluation of mechanical stability in development of customized hip implant," Jouranl of the Korean society of precision engineering, vol. 26, no. 7, pp. 1-2, 2009.
  10. Gudance document for the preparation of premarket notifications of ceramic ball hip systems.
  11. F. Higuchi, N. Shiba, A. Inoue and I. Wakebe, "Fracture of an Alumina Ceramic Head in Total Hip Arthroplasty," The Journal of Arthoplasty, vol. 10, no. 6, 1995.
  12. S. Y. Kim, Y. G. Kim, J. Y. Yeo, D. H. Kim, and J. C. Ihn "Ceramic-on-Ceramic Bearing Total Hip Athroplasty in Young Patients with Osteonecrosis of Femoral Head," J.of Korean Orthop. Assoc. vol. 39, pp. 233-8, 2004. https://doi.org/10.4055/jkoa.2004.39.3.233
  13. A, Arenas, A. Tejero, A. Garbayo, C. Arias, J. Barbadillo, and J. Lecumberri "Ceramic femoral heal fractures I total hip replacement," International Orthopaedics, vol. 23, pp. 351-352, 1999. https://doi.org/10.1007/s002640050390