AN EVALUATION OF ROTATIONAL STABILITY IN ENDODONTIC ELECTRONIC MOTORS

근관치료용 전기모터의 회전 안정성 평가

  • Park, Se-Hee (Department of Conservative Dentistry, College of Dentistry, Gangneung-Wonju National University) ;
  • Seo, Hyun-Woo (Department of Pediatric Dentistry, College of Dentistry, Gangneung-Wonju National) ;
  • Hong, Chan-Ui (Department of Conservative Dentistry, Plant Dental Hospital)
  • 박세희 (강릉원주대학교 치과대학 치과보존학교실) ;
  • 서현우 (강릉원주대학교 치과대학 소아청소년치과학교실) ;
  • 홍찬의 (플랜트 치과병원 치과보존과)
  • Received : 2010.04.28
  • Accepted : 2010.06.11
  • Published : 2010.07.30

Abstract

The purpose of this study was to evaluate a rotational stability of endodontic electronic motors by comparing the changes of rotational speed, depending on the number of usages and with/without static load. Twelve new endodontic electronic motors were used in this study. Non contact type digital tachometer was used for measuring the rotational speed of handpiece. True RMS Multimeter was used for measuring the voltages and the electric currents. All measurements were recorded every 10 seconds during 10 minutes and repeated 9 times. Five repetition was done per each electronic motor. To statistical analysis, student t-test, repeated measures and Scheffe's post-hoc tests were performed. In the same motor group, there was no significant difference in all measurements. In all groups, there was no significant difference in the amount of rotational speed changes depending on the number of usages and with/without static load. In the limitation of this study, the results showed that all kinds of endodontic electronic motors in this study had an established rotational stability. Therefore they could be safely used in root canal treatment with a reliable maintenance of rotational speed, regardless of the number of usages and with/without load.

수종의 근관치료용 전기모터들의 반복사용과 시간에 따른 회전속도의 변화를 비교하여 근관치료용 전기모터의 회전 안정성을 평가하고자 하였다. 시간에 따라 변화하는 실제회전속도 및 전압과 전류를 측정하고, Student t-test를 통한 실제회전속도와, 반복측정분석을 통한 상관관계를 비교분석 하였다. 동일 종류의 전기모터들 사이의 모든 측정 결과들은 차이를 보이지 않았다. 모든 전기모터에서 반복사용과 시간의 흐름에 따른 회전속도의 변화는 차이를 보이지 않았다. 이상의 결과로, 본 실험에 사용된 모든 전기모터는 일정한 회전속도를 유지해야 하는 회전 안정성의 조건에 충족됨을 알 수 있으며, 반복사용과 시간에 따른 신뢰성 있는 회전속도와 안정적인 전압 및 전류의 유지가 가능하리라 생각된다.

Keywords

References

  1. Schilder. Cleaning and shaping the root canal. Dent Clin North Am 18:269-296, 1974.
  2. Skidmore AE, Bjorndal AM, Root canal morphology of human mandibular first molar. Oral Surg Oral Med Oral Pathol 32:778-784, 1971. https://doi.org/10.1016/0030-4220(71)90304-5
  3. Lim KC, Webber J. The effect of root canal preparation on the shape of the curved root canal. Int Endod J 18:233-239, 1985. https://doi.org/10.1111/j.1365-2591.1985.tb00449.x
  4. Kesseler JR, Peters DD, Lorton L. Comparison of the relative risk of molar root perforations using various endodontic instrumentation techniques. J Endod 9:437-439, 1983.
  5. Weine F, Kelly R, Lio P. The effect of preparation procedures on the original canal shape and on apical foramen shape. J Endod 1:255-262, 1975. https://doi.org/10.1016/S0099-2399(75)80037-9
  6. Walia H, Brantely WA, Gerstein H. An initial investigation of the bending and torsional properties of nitinol root canal files. J Endod 14:346-35, 19881. https://doi.org/10.1016/S0099-2399(88)80196-1
  7. Glosson CR, Haller RH, Dove SB, del Rio CE. A comparison of root canal preparations using Ni-Ti hand, Ni-Ti engine-driven, and K-flex endodontic instruments. J Endod 21:146-151, 1995. https://doi.org/10.1016/S0099-2399(06)80441-3
  8. Serene TP, Adams JD, Saxena A. Nickel-titanium instruments: applications in endodontics. St. Louis, Mosby. 67-78, 1995.
  9. Averbach RE, Kleier DJ. Endodontics in the 21st century: the rotary revolution. Compend Contin Educ Dent 22:27-34, 2001.
  10. Spili P, Parashos P, Messer HH. The impact of instrument fracture on outcome of endodontic treatment. J Endod 31:845-50, 2005. https://doi.org/10.1097/01.don.0000164127.62864.7c
  11. Parashos P, Gordon I, Messer HH. Factors influencing defects of rotary nickel-titanium endodontic instruments after clinical use. J Endod 30:722-725, 2004. https://doi.org/10.1097/01.DON.0000129963.42882.C9
  12. Schilder H. Filling root canals in three dimensions. Dent Clin North Am 11:723-744, 1967.
  13. Crump MC, Natkin E. Relationship of broken root canal instruments to endodontic case prognosis: a clinical investigation. J Am Dent Assoc 80:1341-1347, 1970. https://doi.org/10.14219/jada.archive.1970.0259
  14. Hulsmann M, Schinkel I. Influence of several factors on the success or failure of removal of fractured instruments from root canals. Endod Dent Traumatol 15: 252-258, 1999. https://doi.org/10.1111/j.1600-9657.1999.tb00783.x
  15. Souter NJ, Messer HH. Complications associated with fractured file removal using an ultrasonic technique. J Endod 31:450-452, 2005. https://doi.org/10.1097/01.don.0000148148.98255.15
  16. Sattapan B, Nervo GJ, Palamara JEA, Messer HH. Defect in rotary nickel-titanium files after clinical use. J Endod 26:161-165, 2000. https://doi.org/10.1097/00004770-200003000-00008
  17. Cohen S, Burns RC. Pathways of the pulp. 8th ed. St. Louis, Mosby. 533-534, 2002.
  18. Yared GM, Bou Dagher FE, Kulkarni GK. Influence of torque control motors and the operator’s proficiency on ProTaper failures. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 96:229-233, 2003. https://doi.org/10.1016/S1079-2104(03)00167-7
  19. Gabel WP, Hoen M, Steiman HR, Pink FE, Dietz R. Effect of rotational speed on nickel-titanium file distortion. J Endod 25:752-754, 1999. https://doi.org/10.1016/S0099-2399(99)80124-1
  20. Seo MC, Jeon YJ, In-Chol Kang, Kim DJ, Hwang YC, Hwang IN, Oh WM. Effect of rotational speed of ProtaperTM rotary file on the change of root canal configuration. J Kor Acad Cons Dent 31:179-184, 2006. https://doi.org/10.5395/JKACD.2006.31.3.179
  21. Dietz DB, Di Fiore PM, Bahcall JK, Lautenschlager EP. Effect of rotational speed on the breakage of nickel- titanium rotary files. J Endod 26:68-71, 2000. https://doi.org/10.1097/00004770-200002000-00002
  22. Shin JH, Baek SH, Bae KS, Lim SS, Yoon SH, Kim BH. A study of working efficiency and file deformation of GT rotary file in curved canals. J Kor Acad Cons Dent 26:418-435, 2001.
  23. Puolsen BW, Dove BS, Del Rio CE. Effect of nickeltitanium engine-driven instrument rotational speed on root canal morphology. J Endod 21:609-612, 1995. https://doi.org/10.1016/S0099-2399(06)81113-1
  24. Yared G, Kulkarni GK. Accuracy of the DTC torque control motor for nickel-titanium rotary instruments. Int Endod J 34:399-402, 2004.
  25. Yared G, Kulkarni GK. Accuracy of the Nouvag torque control motor for nickel-titanium rotary instruments. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 97:499-501, 2004. https://doi.org/10.1016/j.tripleo.2003.12.021
  26. Yared G, Kulkarni GK. Accuracy of the TCM Endo III torque-control motor for nickel-titanium rotary instruments. J Endod 30:644-645, 2004. https://doi.org/10.1097/01.DON.0000121615.57088.2D
  27. Operation manual of Endomate TC. NSK 10, 2005.
  28. ADA Professional Product Review. Rotary Endodontic Instruments: Laboratory Testing Methods 2006. in American National Standard/American Dental Association Specification No. 28 for root canal files and reamers, type K for hand use. Chicago: American Dental Association, 2002.
  29. Mandel E, Adib-Yazdi M, Benhamou LM, Lachkar T, Mesgouez C, Sobel M. Rotary Ni-Ti ProFile systems for preparing curved canals in resin blocks: influence of operator on instrument breakage. Int Endod J 32:436-443, 1999. https://doi.org/10.1046/j.1365-2591.1999.00239.x
  30. Yared GM, BouDagher FE, Machtou P. Influence of rotational speed, torque, and operator’s proficiency on ProFile failures. Int Endod J 34:47-53, 2001. https://doi.org/10.1046/j.1365-2591.2001.00352.x
  31. Yared GM, Bou Dagher FE, Machtou P, Kulkarni GK. Influence of rotational speed, torque and operator proficiency on failure of Greater Taper files. Int Endod J 35:7-12, 2002. https://doi.org/10.1046/j.1365-2591.2002.00443.x
  32. Gambarini G. Cyclic fatigue of nickel-titanium rotary instruments after clinical use with low- and hightorque endodontic motors. J Endod 27:772-774, 2001. https://doi.org/10.1097/00004770-200112000-00015
  33. Yared G, Sleiman P. Failure of ProFile instruments with air, high torque control and low torque control motors. Int Endod J 34:471-475, 2001. https://doi.org/10.1046/j.1365-2591.2001.00420.x
  34. Pruett JP, Clement DJ, Carnes DL Jr. Cyclic fatigue testing of nickel titanium endodontic instruments. J Endod 23:77-85, 1997. https://doi.org/10.1016/S0099-2399(97)80250-6
  35. Gambarini G. Cyclic fatigue of ProFile rotary instruments after prolonged clinical use. Int Endod J 34: 386-389, 2001. https://doi.org/10.1046/j.1365-2591.2001.00259.x
  36. Yared GM, Bou Dagher FE, Machtou P. Cyclic fatigue of Profile rotary instruments after simulated clinical use. Int Endod J 32:115-119, 1999. https://doi.org/10.1046/j.1365-2591.1999.00201.x
  37. Yared GM, Bou Dagher FE, Machtou P. Cyclic fatigue of Profile rotary instruments after clinical use. Int Endod J 33:204-207, 2000. https://doi.org/10.1046/j.1365-2591.1999.00296.x
  38. Shin YM, Kim ES, Kim KM, Kum KY. Effect of surface defects and cross-sectional configuration on the fatigue fracture of NiTi rotary files under cycllc loading. J Kor Acad Cons Dent 29:267-272, 2004. https://doi.org/10.5395/JKACD.2004.29.3.267
  39. Kim JW, Ahn BD, Park SH, Shin HJ, Cho KM. Lifetime and fracture patterns of NiTi rotary files in molars. J Kor Acad Cons Dent 30:190-198, 2005.
  40. Blum JY, Machtou P, Micallef JP. Location of contact areas on rotary Profile instruments in relationship to the forces developed during mechanical preparation on extracted teeth. Int Endod J 32:108-114, 1999. https://doi.org/10.1046/j.1365-2591.1999.00200.x
  41. Thompson SA, Dummer PM. Shaping ability of Quantec Series 2000 rotary nickel-titanium instruments in simulated root canals. Part 1. Int Endod J 31:259-267, 1998. https://doi.org/10.1046/j.1365-2591.1998.00151.x
  42. Thompson SA, Dummer PM. Shaping ability of ProFile .04 Taper Series 29 rotary nickel-titanium instruments in simulated root canals. Part 2. Int Endod J 30:8-15, 1997. https://doi.org/10.1111/j.1365-2591.1997.tb01092.x
  43. Barbakow F, Lutz F. The Lightspeed preparation technique evaluated by Swiss clinicians after attending continuing education courses. Int Endod J 30:46-50, 1997. https://doi.org/10.1111/j.1365-2591.1997.tb01097.x
  44. Yu JS, Han SC, Lee JY. The surface modification of Ti-based alloy with long cycle life for Ni/MH secondary battery. Trans. of the Korean Hydrogen and New Energy Society 11:151-161, 2000.
  45. Park CN. Characteristics and Research Trends of Ni/MH secondary battery. Bulletin of KIEEME 12:28-33, 1991.
  46. Berutti E, Negro AR, Lendini M, Pasqualini D. Influence of manual preflaring and torque on the failure rate of ProTaper rotary instruments. J Endod 30: 228-230, 2004. https://doi.org/10.1097/00004770-200404000-00011
  47. Bahia MGA, Martins RC, Gonzalez BM, Buono VTL. Physical and mechanical characterization and the influence of cyclic loading on the behavior of nickeltitanium wires employed in the manufacture of rotary endodontic instruments. Int Endod J 38:795-801, 2005. https://doi.org/10.1111/j.1365-2591.2005.01016.x
  48. Kobayashi C, Yoshioka T, Suda H. A new engine-driven canal preparation system with electronic canal measuring capability. J Endod 23:751-754, 1997. https://doi.org/10.1016/S0099-2399(97)80349-4
  49. Li UM, Lee BS, Shih CT, Lan WH, Lin CP. Cyclic fatigue of endodontic nickel-titanium rotary instruments: static and dynamic tests. J Endod 28:448-451, 2002. https://doi.org/10.1097/00004770-200206000-00007
  50. Roane J, Sabala C, Duncanson M. The balanced force concept for instrumentation of curved canals. J Endod 11:203-211, 1985. https://doi.org/10.1016/S0099-2399(85)80061-3
  51. Kyomen SM, Caputo AA, White SN. Critical analysis of the balanced force technique in endodontics. J Endod 20:332-337, 1994. https://doi.org/10.1016/S0099-2399(06)80095-6
  52. Petka K, Implementing one-visit root canal therapy. Dent Today 17:112-117, 1998.

Cited by

  1. Effect of internal stress on cyclic fatigue failure in K3 vol.37, pp.2, 2012, https://doi.org/10.5395/rde.2012.37.2.74
  2. Effect of internal stress on cyclic fatigue failure in .06 taper ProFile vol.37, pp.2, 2012, https://doi.org/10.5395/rde.2012.37.2.79
  3. Rotational stability of endodontic motors vol.39, pp.3, 2014, https://doi.org/10.5395/rde.2014.39.3.240