A Comparison of Lens Centrations on Cornea with RGP Lens Fitting by the Measured Values using Keratometer and Corneal Topography

각막곡률계와 각막 지형도 검사에서의 측정값을 이용한 RGP 렌즈 피팅시 각막에서의 중심안정위치 비교

  • Kim, So Ra (Department of Visual Optics, Seoul National University of Science and Technology) ;
  • Park, Sang-Il (Department of Visual Optics, Seoul National University of Science and Technology) ;
  • Lee, Se Eun (Department of Visual Optics, Seoul National University of Science and Technology) ;
  • Park, Mijung (Department of Visual Optics, Seoul National University of Science and Technology)
  • 김소라 (서울과학기술대학교 안경광학과) ;
  • 박상일 (서울과학기술대학교 안경광학과) ;
  • 이세은 (서울과학기술대학교 안경광학과) ;
  • 박미정 (서울과학기술대학교 안경광학과)
  • Received : 2011.01.29
  • Accepted : 2011.03.19
  • Published : 2011.03.31

Abstract

Purpose: The present study was conducted to compare the centration of RGP lens on cornea when lens was fitted based on keratometric astigmatisms measured by keratometer and the lens centration when fitted by corneal topography. Methods: Thirty eight eyes of 19 male and female in their twenties were applied RGP lens with 9.9 mm of diameter by the keratometric astigmatisms classified by the measurement with a keratometer. Then, lens centrations were estimated using high speed camera and compared with the lens centration when fitted by total keratometric astigmatism using corneal topography. The relationship of the steepest location of cornea and lens centration was further compared. Results: With the rule astigmatism, lens centration was not changed even with the difference in central and total keratometric astigmatisms. When the relationship of the steepest part of cornea measured by corneal topography and lens centration was analyzed, the lens centration in vertical direction was exactly correlated with the steepest part of cornea in 52.3% of subjects. In the case of non-correlation, the steepest part of cornea was mostly upper part of cornea, however, lens centration was located on lower part of cornea. The lens centration in horizontal direction was exactly correlated with the steepest region of cornea in 65.6% of subjects. In non-correlated case, the difference in cornea curvatures between the steepest and the flattest parts was smaller than 0.05 mm in 76.9% of subjects. Conclusions: From these results, we conclude that corneal topographic patterns may more contribute the centration of RGP lens on cornea than the difference in central and total keratometric astigmatisms.

목적: 각막곡률계(keratometer) 측정값을 기준으로 분류된 난시량에 의해 피팅(fitting)하였을 때 RGP 렌즈의 중심 안정위치가 각막 지형도 검사(corneal topography) 측정값을 기준으로 각막 난시를 분류하여 전체 각막 난시를 고려하였을 때 어떻게 달라지는지를 알아보고자 하였다. 방법: 직난시를 가진 20~30대 남녀 19명 38안에게 각막곡률계 측정값으로 분류한 각막 난시량에 따라 9.9 mm 직경의 비구면 RGP 렌즈를 피팅한 후 초고속 촬영장치를 이용하여 각막에서 렌즈의 중심안정위치를 측정하였다. 또한 각막 지형도 검사에 의해 측정된 전체 각막 난시량을 기준으로 하였을 때와 각막의 가장 가파른 부분의 위치와 렌즈의 중심안정위치 관계를 비교하였다. 결과: 직난시안에서는 중심부 각막 난시값과 각막 전체 난시값의 차이가 있을 때에 RGP 렌즈의 중심 안정 위치에는 변화가 없었다. 각막지형도 검사를 이용하여 각막에서 가장 가파른(steep) 위치와 RGP 렌즈의 중심안정위치와의 관계를 알아보았을 때 각막에서 가장 가파른 위치와 RGP 렌즈의 수직방향에서의 중심안정위치가 일치하는 경우는 52.3%였으며, 두 위치가 일치하지 않는 경우의 대부분은 각막의 가파른 부분이 윗방향인데 반하여 렌즈의 중심안정위치는 아랫방향인 경우였다. 각막의 가장 가파른 위치와 렌즈의 수평방향에서의 중심안정위치가 일치하는 경우는 65.6%였으며, 이들 두 위치가 일치하지 않는 경우의 76.9%는 각막의 가장 가파른 부분의 곡률반경과 플랫(flat)한 부분의 곡률반경 차이가 0.05 mm 이하로 작았다. 결론: 본 연구에서는 RGP 렌즈의 중심안정위치는 중심부 각막 난시와 전체 각막 난시의 차이에 의해 영향을 받지 않으며 오히려 각막의 지형이 더 중요할 수 있다는 것을 밝혔다.

Keywords

References

  1. Albright R. A., Venuti B. D., Ichijima H., Nyunt A. K., and Cavanagh H. D., "Postmarket surveillance of Menicon Z rigid gas-permeable contact lenses for up to 30 days continuous wear in the United States", Eye Contact Lens, 36(5):241-244(2010). https://doi.org/10.1097/ICL.0b013e3181efa61b
  2. Martin R., "Corneal conjunctivalization management with high Dk RGP contact lenses", Cont. Lens Anterior Eye, 32(3):147-150(2009). https://doi.org/10.1016/j.clae.2008.12.005
  3. Szczotka-Flynn L. B. and Patel S., "Menicon Z rigid gas permeable lenses for keratoconus and irregular corneas: a retrospective case series", Eye Contact Lens, 34(5):254-260(2008). https://doi.org/10.1097/ICL.0b013e31817f6db0
  4. Nosch D. S., Ong G. L., Mavrikakis I., and Morris J., "The application of a computerised videokeratography (CVK) based contact lens fitting software programme on irregularly shaped corneal surfaces", Cont. Lens Anterior Eye, 30(4):239-248(2007). https://doi.org/10.1016/j.clae.2007.06.003
  5. Efron N., "Obituary rigid contact lenses", Cont. Lens Anterior Eye, 33(5):245-252(2010). https://doi.org/10.1016/j.clae.2010.06.009
  6. Davis R., Keech P., Dubow B., and Ames K., "Making RGP fitting efficient and successful", Contact Lens Spectrum, 15(10):40-46 (2000).
  7. Kok J. H., "A European fitting philosophy for aspheric, high-Dk RGP contact lenses", CLAO J., 18(4):232-236(1992).
  8. 김재호, "RGP 콘택트렌즈", 초판, 현문사, 서울, pp.69-81(1998).
  9. Ming W., Tracy S., Renzo M., Nancy K. T., and Doug H., "Corneal topography in wavefront era: A guide for clinical application", Graefe's archive for clinical and experimental ophthalmology, 246(2):327-328(2006).
  10. Clark B. A. and Knoll H. A., "Variations in corneal topography", Optom. Vis. Sci., 56:399-413(1983).
  11. Faber E., Golding T. R., Lowe R., and Brennan N. A., "Effect of hydrogel lens wear on tear film stability", Optom. Vis. Sci., 68:380-384 (1991). https://doi.org/10.1097/00006324-199105000-00010
  12. 박미정, 양재헌, 김선미, 박상일, 박상희, 김소라, "일일착용렌즈의 착용기간 초과 시 유발되는 렌즈의 중심안정위치 및 순목횟수 변화", 한국안광학회지, 13(4):51-58 (2008).
  13. 박상일, 이연진, 이흠숙, 박미정, "순목에 의한 소프트렌즈의 순간적인 움직임; 착용시간의 증가에 따른 움직임 변화", 한국안광학회지, 12(1):1-7 (2007).
  14. 최병철, 박성현, "SPSS와 SAS분석을 통한 실험계획법의 이해", 초판, 민영사, 서울, pp. 26-30(2005).
  15. 박미라, 이재원, "EXCEL을 이용한 의학데이터의 통계분석", 개정판, 자유 아카데미, pp.159-165(2007).
  16. Bogan S. J., Waring G. O., Ibrahim O., and Drews C., "Classification of normal corneal topography based on computer-assisted videokeratography", Arch. Ophthalmol., 108:945-949(1990). https://doi.org/10.1001/archopht.1990.01070090047037
  17. 김대현, 마기중, 김대수, 이혜정, "순목속도에 따른 RGP 콘택트렌즈의 움직임", 대한시과학회지, 9(2):225-239(2007).
  18. 박영진, 변장원, 마기중, 박수봉, 원찬희, "자동각막곡률계와 Corneal topography System의 비교", 대한시과학회지, 1(1):111-123(1999).
  19. Theodore B., Anastasios G. P. K., and Effrosini M., "The role of computerized corneal topography in rigid gas permeable contact lens fiting", CLAO J, 24(4):206-209(1998).
  20. Worp E. V. D., Brabander J., Lubberman B., and Hendriske F., "Optimising RGP lens fitting in normal eyes using 3D topographic data", Cont. Lens Anterior Eye, 25(2):95-99(2002). https://doi.org/10.1016/S1367-0484(02)00006-1