DOI QR코드

DOI QR Code

Study of The Amorphous Selenium (a-Se) using 2-dimensional Device Simulator

2차원 소자 시뮬레이터를 이용한 비정질 셀레늄(a-Se) 분석

  • Kim, Si-Hyoung (School of Electronic Engineering, Kumoh National Institute of Technology) ;
  • Kim, Chang-Man (Biomedical IT Convergence Center, Gumi Electronics & Information Technology Research Institute) ;
  • Nam, Ki-Chang (Department of Medical Engineering, Yonsei University College of Medicine & Severance MDCTC) ;
  • Kim, Sang-Hee (School of Electronic Engineering, Kumoh National Institute of Technology) ;
  • Song, Kwang-Soup (Department of Medical IT Convergence Engineering, Kumoh National Institute of Technology)
  • 김시형 (금오공과대학교 전자공학부) ;
  • 김창만 (구미전자정보기술원 IT의료융합기술사업단) ;
  • 남기창 (연세대학교 의과대학 의학공학연구실, 세브란스 의료기기임상시험센터) ;
  • 김상희 (금오공과대학교 전자공학부) ;
  • 송광섭 (금오공과대학교 메디컬IT융합공학과)
  • Received : 2012.08.17
  • Published : 2012.10.25

Abstract

Digital X-ray image detector has been applied for medical and industrial fields. Photoconductors have been used to convert the X-ray energy to electrical signal on the direct digital X-ray image detector and amorphous selenium (a-Se) has been used as a photoconductor, normally. In this work, we use 2-dimensional device (2-D) simulator to study about physical phenomena in the a-Se, when we irradiate electromagnetic radiation (${\lambda}=486nm$) on the a-Se surface. We evaluate the electron-hole generation rate, electron-hole recombination rate, and electron/hole distribution in the a-Se using 2-D simulator. This simulator divides the device into triangle and calculates using interpolation method. This simulation method has been proposed for the first time and we expect that it will be applied for the development of digital X-ray image detector.

디지털 X-ray 영상 디텍터는 의료용 및 산업용으로 널리 이용되고 있다. 직접방식(direct method)의 디지털 X-ray 영상 디텍터는 X-ray 에너지를 전기적 신호로 변환하기 위하여 광도전체(photoconductor)를 이용하며 일반적으로 비정질 셀레늄(a-Se)을 사용하고 있다. 본 연구는 비정질 셀레늄 표면에 파장 486 nm의 전자방사선을 조사할 경우 내부에서 일어나는 물리적 현상들을 분석하기 위하여 2차원 소자 시뮬레이터을 이용하였다. 시뮬레이션 결과를 통해 비정질 셀레늄 내부 전자-정공 생성율, 전자-정공 재결합율, 전자/정공 분포에 대한 분석을 수행하였다. 사용된 시뮬레이터는 디바이스 내부를 삼각형으로 나누어 보간법을 사용하여 계산하는 방식이다. 본 시뮬레이션 방법은 직접방식 디지털 X-ray 영상 디텍터 분석을 위하여 처음으로 제안되었고 유용한 방법이다. 이러한 2차원 소자 시뮬레이터를 이용한 연구방법은 향후 디지털 방사선 영상 디텍터 개발에 많이 응용될 것으로 예상된다.

Keywords

References

  1. Wei Zhaoa, D.C. Huntb, Kenkichi Taniokac, J.A. Rowlandsb, "Amorphous selenium flat panel detectors for medical applications", Physics Research A, 549, 205-209, 2005.
  2. Jang Yong Choi, Ji Koon Park, Sang Sik Kang, Chi Woong Moon, Hyung Won Lee and Sang Hee Nam, "비정질 셀레늄을 이용한 직접방식의 디지털 방사선 검출기와 X-ray film과의 MTF 측정을 통한 영상 질(Quality) 비교 평가에 관한연구", 대한전자공학회, 40, 71-79, 2003.
  3. 차보경, 이동훈, 김봉회, 황윤금, 전성채, 허영, "디지털 3D 투시 X-ray 영상 센서용 섬광체 연구개발", 대한전자공학회 2010년 하계학술대회, 33, 2025-2027, 2010.
  4. S.O. Kasap, John Rowlands, B. Fogal, M. Zahangir Kabir, G. Belev, N. Sidhu, Brad Polischuk, Robert E. Johanson, "Progress in the science and technology of direct conversion a-Se X-ray sensors", Journal of Non-Crystalline Solids, 299-302, 988-992, 2002. https://doi.org/10.1016/S0022-3093(01)01064-X
  5. J.A. Rowlands, J. Yorkston, in: J. Beutel, H.L. Kundel, R.L. Van Metter (Eds.), Handbook of Medical Imaging, vol. 1, SPIE Press, Washington, Chapter 4 and references therein, 2000.
  6. Martin Hoheisel, Lothar Batz, "Requirements on amorphous semiconductors for medical X-ray detector", Thin Solid Films, 383, 130-136, 2001.
  7. G. Belev, S.O. Kasap, "Amorphous selenium as X-ray photoconductor", Journal of Non-Crystalline Solids, 345&346, 484-488, 2004. https://doi.org/10.1016/j.jnoncrysol.2004.08.070
  8. Shiva Abbaszadeh, Nicholas Allec, Kai Wang, Karim S. Karim, "Low Dark-Current Lateral Amorphous-Selenium Metal-Semiconductor- Metal Photodetector", IEEE ELECTRON DEVICE LETTERS, 32, 1263-1265, 2011. https://doi.org/10.1109/LED.2011.2160327
  9. Ho Kyung Kim, Min Kook Cho, Seong Kyu Ahn, Gyu seong Cho and Hyo sung Cho, "Electrical Investigation of LIGA-Processed Gas Electron Multipliers", Journal of the Korean Physical Society, 48, 40-46, 2006.
  10. http://www.silvaco.com/tech_lib_TCAD/simulation standard/2011/oct_nov_dec/a2/a2.html
  11. ATLAS User's Manual, 2012.
  12. Kai Wang, Feng Chen, Nicholas Allec and Karim S., "Fast Lateral Amorphous-Selenium Metal-Semiconductor-Metal Photodetector with High Blue-to-Ultraviolet Responsivity", IEEE, 57, 8, 1953-1958, 2010.
  13. Kai Wang, Feng Chen, Nicholas Allec, Fang Yuan, George Belev, Safa Kasap and Karim S. Karim, "Amorphous-Selenium-Based Three- Terminal X-ray Detector With a Gate", IEEE, 32, 782-784, 2011.
  14. W. Que and J. A. Rowlands, "X-ray photogeneration in amorphous selenium: Geminate versus columnar recombination", Phys. Rev. 51, 500-507, 1995. https://doi.org/10.1103/PhysRevA.51.500

Cited by

  1. Design and Fabrication of HgI2 Sensor for Phosphor Screen based flat panel X-ray Detector vol.51, pp.12, 2014, https://doi.org/10.5573/ieie.2014.51.12.189