DOI QR코드

DOI QR Code

Fabrication and Evaluation of Sensor for Measuring Pulse Wave Velocity using Piezo Film and Conductive Textile

압전 필름과 전도성 섬유를 이용한 맥파 전달 속도 측정을 위한 센서의 제작 및 성능평가

  • Kim, Jung-Chae (Graduate School of Biomedial Engineering, Yonsei University) ;
  • Jee, Sun-Ha (Department of Epidemiology and Health Promotion, Graduate School of Public Health, Yonsei University) ;
  • Yoo, Sun-Kook (Department of Medical Engineering, College of Medicine, Yonsei University)
  • 김정채 (연세대학교 일반대학원 생체공학협동과정) ;
  • 지선하 (연세대학교 보건대학원 역학건강증진학과) ;
  • 유선국 (연세대학교 의과대학 의학공학교실)
  • Received : 2011.11.05
  • Accepted : 2012.02.04
  • Published : 2012.03.31

Abstract

Arterial stiffness is causing the serious problems for human who is suffered from hypertension and metabolic syndrome. So it is important that measure the arterial stiffness for early prevention. Many researches point out that pulse wave velocity(PWV) is the reliable and simple method to predict arterial stiffness. In this paper, we developed the sensing parts that detect the pulse wave and ECG by using piezoelectric film and conductive textile with elastic band. Our system could detect 3ch pulse wave and ECG. Simultaneously, our algorithm extracts the features for calculating the delays among pulse waves. The delays are the significant parameter to estimate PWV, thus we design the experiment for evaluating the performance of our sensing parts. The reference is PP-1000(HanByul Meditech, Korea) that is good for performance evaluation. As a result, the start point of the pulse wave was the most reliable feature for comparing with PP-1000(r=0.691, P=0.00). The results between two operators showed that there is only a slight difference in the reproducibility of the devices. In conclusion, we assume that the suggested sensor could be more comfortable and faithful method for arterial stiffness.

Keywords

References

  1. Safar ME, Levy BI, Laurent S, and London GM, "Hypertension and the arterial system: Clinical and therapeutic aspects", J. Hypertension, vol. 8, no. 7, pp. 113-119, 1990.
  2. Bader H, "Importance of the gerontology of elastic arteries in the development of essential hypertension", Clin Physiol Biochem., no. 1, pp. 36- 56, 1983.
  3. Frank O, "Die grundform des arteriellen pulses", Z Biol., vol. 37, pp. 483-526, 1899.
  4. Johnson CP, Baugh R, Wilson CA, and Burns J, "Age related changes in the tunica media of the vertebral artery: implications for the assessment of vessels injured by trauma", J. Clin Pathol., vol. 54, pp. 139-45, 2001. https://doi.org/10.1136/jcp.54.2.139
  5. Hanna-Maaria Lakka, David E. Laaksonen, Timo A. Lakka, Leo K. Niskanen, Esko Kumpusalo, Jaakko Tuomilehto, and Jukka T. Salonen, "The metabolic syndrome and total and cardiovascular disease mortality in middle-aged men", The Journal of the American Medical Association, vol. 288, no. 22, pp. 2709-2716, 2002. https://doi.org/10.1001/jama.288.21.2709
  6. C. D. A. Stehouwer, R. M. A. Henry, and I. Ferreira, "Arterial stiffness in diabetes and the metabolic syndrome: a pathway to cardiovascular disease", Diabetologia, vol. 51, no. 4, pp. 527-539, 2008. https://doi.org/10.1007/s00125-007-0918-3
  7. Stephane Laurent, John Cockcroft, Luc Van Bortel, Pierre Boutouyrie, Cristina Giannattasio, Daniel Hayoz, Bruno Pannier, Charalambos Vlachopoulos, Ian Wilkinson, and Harry Struijker-Boudier, "Expert consensus document on arterial stiffness: methodological issues and clinical applications", European Heart Journal, vol. 27, no. 21, pp. 2588- 2605, 2006. https://doi.org/10.1093/eurheartj/ehl254
  8. Measurement Specialties 2006 Piezo Film Sensor Technical Manual (Hampton, VA: Measurement Specialties Inc.).
  9. Gruetzmann Anna, Hansen Stefan, and Muller Jorg, "Novel dry electrodes for ECG monitoring", Physiological Measurement, vol. 28, no. 11, pp. 1375-1390, 2007. https://doi.org/10.1088/0967-3334/28/11/005
  10. Bert-Uwe Kohler, Carsten Hennig, and Reinhold Orglmeister, "The principles of software QRS detection", IEEE Engineering in Medicine and Biology Magazine, vol. 22, no. 1, pp. 42-57, 2002.
  11. Y. Christopher Chiu, Patricia W. Arand, Sanjeev G. Shroff, Ted Feldman, and John D. Carroll, "Determination of pulse wave velocities with computerized algorithms", American Heart Journal, vol. 121, no. 5, pp. 1460-1470, 1991. https://doi.org/10.1016/0002-8703(91)90153-9
  12. Masanori Munakata, Nobuhiko Ito, Tohru Nunokawa, and Kaoru Yoshinaga, "Utility of automated brachial ankle pulse wave velocity measurements in hypertensive patients", American Journal of Hypertension, vol. 16, no. 8, pp. 653- 657, 2003. https://doi.org/10.1016/S0895-7061(03)00918-X
  13. Roland Asmar, Athanase Benetos, Jirar Topouchian, Pierre Laurent, Bruno Pannier, Anne- Marie Brisac, Ralph Target, and Bernard I. Levy, "Assessment of arterial distensibility by automatic pulse wave velocity measurement", Hypertensions, vol. 26, no. 3, pp. 485-490, 1995. https://doi.org/10.1161/01.HYP.26.3.485
  14. Park, Jong Suk, Nam, Ji Sun, Cho, Min Ho, Yoo, Jeong Seon, Ahn, Chul Woo, Jee, Sun Ha, Lee, Hong Soo, Cha, Bong Soo, Kim, Kyung Rae, Lee, and Hyun Chul, "Insulin resistance independently influences arterial stiffness in normoglycemic normotensive postmenopausal women", Menopause, vol. 17, no. 4, pp. 779-784, 2010.
  15. J.S. Nam, J.S. Park, M.H. Cho, S.H. Jee, H.S. Lee, C.W. Ahn, W.L. Lowe, and K.R. Kim, "The association between pulse wave velocity and metabolic syndrome and adiponectin in patients with impaired fasting glucose: Cardiovascular risks and adiponectin in IFG", vol. 84, no. 2, pp. 145-151, 2009. https://doi.org/10.1016/j.diabres.2009.02.002
  16. C. Fabrizio, A. Pasquale, and C. Pasquale. "A piezo-film-based measurement system for global haemodynamic assessment", Physiological Measurement, vol. 31, no. 5, pp. 697-714, 2010. https://doi.org/10.1088/0967-3334/31/5/007
  17. Junghyeon Choi, Wookhyun Cho, Junho Park, Namhoon Kim, Hyangsook Seong, and Jongman Cho, "Development of feature points detection algorithm for measuring of pulse wave velocity", Journal of Sensor Science and Technology, vol. 20, no. 5, pp. 343-350, 2011. https://doi.org/10.5369/JSST.2011.20.5.343
  18. Sang-Joong Jung, Yong-Su Seo, and Wan-Young Chung, "Pulse wave analysis system using wrist type oximeter for u-Health service", J. Kor. Sensors Soc., vol. 19, no. 1, pp. 17-24, 2010. https://doi.org/10.5369/JSST.2010.19.1.017
  19. Nak Bum Lee and Chang Gyu Park, "Reproducibility of regional pulse wave velocity in healthy subjects", Korean Journal of Internal Medicine, vol. 24, no. 1, pp. 19-23, 2009. https://doi.org/10.3904/kjim

Cited by

  1. Wireless Optical Fiber Interferometer Arterial Pulse Wave Sensor System vol.22, pp.6, 2013, https://doi.org/10.5369/JSST.2013.22.6.439
  2. Development of pulse diagnostic devices in Korea vol.2, pp.1, 2013, https://doi.org/10.1016/j.imr.2013.01.003
  3. A Study on Performance Evaluation for Electrocardiography Signal Measurement Electrode based on Conductive Fabric vol.50, pp.2, 2013, https://doi.org/10.5573/ieek.2013.50.2.210
  4. Accuracy Analysis of Pulse Wave Sensor Data of Ear Label of Husbandry Livestock vol.12, pp.11, 2014, https://doi.org/10.14400/JDC.2014.12.11.387