Evaluation of Scattered Dose to the Contralateral Breast by Separating Effect of Medial Tangential Field and Lateral Tangential Field: A Comparison of Common Primary Breast Irradiation Techniques

유방암 접선조사 치료 방법에 대한 반대쪽 유방에서의 산란선량 평가

  • Ban, Tae-Joon (Department of Radiation Oncology, Asan Medical Center) ;
  • Jeon, Soo-Dong (Department of Radiation Oncology, Asan Medical Center) ;
  • Kwak, Jung-Won (Department of Radiation Oncology, Asan Medical Center) ;
  • Baek, Geum-Mun (Department of Radiation Oncology, Asan Medical Center)
  • 반태준 (서울아산병원 방사선종양학과) ;
  • 전수동 (서울아산병원 방사선종양학과) ;
  • 곽정원 (서울아산병원 방사선종양학과) ;
  • 백금문 (서울아산병원 방사선종양학과)
  • Received : 2012.06.18
  • Accepted : 2012.08.24
  • Published : 2012.09.30

Abstract

Purpose: The concern of improving the quality of life and reducing side effects related to cancer treatment has been a subject of interest in recent years with advances in cancer treatment techniques and increasing survival time. This study is an analysis of differing scattered dose to the contralateral breast using common different treatment techniques. Materials and Methods: Eclipse 10.0 (Varian, USA) based $30^{\circ}$ EDW (Enhanced dynamic wedge) plan, $15^{\circ}$ wedge plan, $30^{\circ}$ wedge plan, Open beam plan, FiF (field in field) plan were established using CT image of breast phantom which in our hospital. Each treatment plan were designed to exposure 400 cGy using CL-6EX (VARIAN, USA) and we measured scattered dose at 1 cm, 3 cm, 5 cm, 9 cm away from medial side of the phantom at 1 cm depth using ionization chamber (FC 65G, IBA). We carried out measurement by separating effect of medial tangential field and lateral tangential field and analyze. Results: The evaluation of scattered dose to contralateral breast, $30^{\circ}$ EDW plan, $15^{\circ}$ wedge plan, $30^{\circ}$ wedge plan, Open beam plan, FIF plan showed 6.55%, 4.72%, 2.79%, 2.33%, 1.87% about prescription dose of each treatment plan. The result of scattered dose measurement by separating effect of medial tangential field and lateral tangential field results were 4.94%, 3.33%, 1.55%, 1.17%, 0.77% about prescription dose at medial tangential field and 1.61%, 1.40%, 1.24%, 1.16%, 1.10% at lateral tangential field along with measured distance. Conclusion: In our experiment, FiF treatment technique generates minimum of scattered dose to contralateral breast which come from mainly phantom scatter factor. Whereas $30^{\circ}$ wedge plan generates maximum of scattered doses to contralateral breast and 3.3% of them was scattered from gantry head. The description of treatment planning system showed a loss of precision for a relatively low scatter dose region. Scattered dose out of Treatment radiation field is relatively lower than prescription dose but, in decision of radiation therapy, it cannot be ignored that doses to contralateral breast are related with probability of secondary cancer.

목 적: 암치료 기술의 발전으로 환자의 생존기간이 길어짐에 따라 치료 이후 삶의 질을 증진하고 치료 방법에 의한 부작용을 줄이는 노력에 관심이 집중되고 있다. 본 연구는 유방암 접선조사 치료에서 치료방법 차이가 반대쪽 유방 산란 선량에 미치는 영향을 분석하고자 하였다. 대상 및 방법: 본원에서 제작한 유방 모형 팬텀의 전산화 단층 영상을 이용하여 이클립스 10.0 (Eclipse 10.0, Varian, USA) 치료계획 시스템을 사용하여 $30^{\circ}$ wedge plan, $15^{\circ}$ wedge plan, $30^{\circ}$ EDW (Enhanced dynamic wedge) plan, Non-wedge plan, FIF(Field in Field) plan을 수립하였다. 각 치료계획은 선형가속기 CL-6EX (VARIAN, USA)를 이용하여 400 cGy씩 조사하였고 팬텀의 중심점으로부터 횡축방향으로 1 cm, 3 cm, 5 cm, 9 cm 씩 이동한 지점의 1 cm 깊이에서 전리조(FC 65G, IBA)를 이용하여 내측접선(Medial tangential) 조사와 외측접선(Lateral tangential) 조사에서 발생하는 산란선량을 각각 측정하고 비교 분석하였다. 결 과: 반대쪽 유방 산란 선량을 평가해보았을 때 $30^{\circ}$ wedge plan, $15^{\circ}$ wedge plan, $30^{\circ}$ EDW (Enhanced dynamic wedge) plan, Non-wedge plan, FiF (Field in Field) plan에서 처방선량에 대해 각각 6.55%, 4.72%, 2.79%, 2.33%, 1.87%로 나타났다. 내측접선조사와 외측접선 조사로 나누어 보았을 때 내측접선 조사 측정값은 각각 4.94%, 3.33%, 1.55%, 1.17%, 0.77%로 나타났고 외측접선 조사는 각각 1.61%, 1.40%, 1.24%, 1.16%, 1.10%의 산란 선량이 측정되었다. 결 론: 유방암 접선 조사 치료방법 중 반대쪽 유방에 가장 적은 산란 선량이 발생하는 방법은 FiF plan으로 이때 발생한 산란선량은 팬텀 내에서 기인한 선량이 주로 작용하는 것으로 판단되었다. 가장 많은 산란 선량이 발생하는 치료방법은 $30^{\circ}$ wedge plan이었고 쐐기필터를 비롯한 치료 장비에서 기인한 선량은 3.3%로 평가되었다. 치료계획 시스템은 처방선량에 대해 상대적으로 낮은 산란 선량 영역은 정확성이 떨어지는 것으로 나타났다. 치료 조사야 밖으로 발생하는 산란 선량은 처방 선량에 비해 그 양이 적지만 2차 암 발생 확률과 관련이 있다는 점에서 간과할 수 없는 부분이며 방사선 치료를 결정하는데 있어 고려되어야 할 부분으로 사료된다.