DOI QR코드

DOI QR Code

Dose assessment according to Differences in the Content of Iodine in Contrast Media used in Interventional Procedure

중재적 시술에 사용하는 조영제의 요오드(53I) 함유량 차이에 따른 피폭선량 평가

  • 박혁 (전남대학교병원 영상의학과) ;
  • 전주섭 (전남대학교병원 영상의학과) ;
  • 김용완 (전남대학교병원 영상의학과) ;
  • 장성주 (동신대학교 방사선학과)
  • Received : 2013.12.10
  • Accepted : 2014.03.17
  • Published : 2014.03.28

Abstract

The purpose of this study is to evaluate exposure dose and absorbability according to differences in the content of iodine in contrast media used in interventional procedure and angiography. There were 7 levels of iodine content as 240, 270, 300, 320, 350, 370, and 400 I $mg/m{\ell}$. The higher iodine content was, the more elevated exposure dose and absorbability were. The output dose was about 8.5% from iodine content with 350I $mg/m{\ell}$ higher than that with 270 I $mg/m{\ell}$ in TACE. These results mean that contrast media can have an effect on patient's exposure dose and high iodine content contrast media results in elevation of patient's exposure dose.

중재적 시술 및 혈관조영검사에 사용되는 조영제의 요오드 함유량 차이에 따른 흡수도와 피폭선량을 알아보았다. 실험에 사용된 조영제는 요오드 함유량(I $mg/m{\ell}$)에 따라 240, 270, 300, 320, 350, 370, 400의 7종류를 사용하였으며, 그 결과 요오드 함유량이 많을수록 흡수도 및 피폭선량이 높아짐을 알 수 있었으며, TACE 검사에서도 조영제의 농도 270 I $mg/m{\ell}$에 비해 350 I $mg/m{\ell}$의 출력 선량이 8.5% 정도 증가하였다. 본 연구를 통해 조영제가 환자의 피폭선량에 영향을 주며 요오드 함유량이 많은 조영제일수록 환자의 피폭선량이 증가함을 알 수 있었다.

Keywords

References

  1. 도영수, 혈관조영 X-선 장치의 안전관리 기준 개발, 식품의약품안전청 용역연구개발사업, 2003.
  2. ICRP Publication 103 : Recommendations of the International Commission on Radiological Protection, Oxford, 2007.
  3. 정진욱, 인터벤션영상의학, 대한인터벤션영상의학회, 2007.
  4. 이재기, 방사선 방호를 위해 표준한국인 설정 연구, 한양대학교, 2004.
  5. http://www.healthcare.philips.com/pwc_hc/main/shared/Assets/Documents/DoseWise/allura_xper_cardiac_system_imp_auto_dose_rate_ctrl.pdf
  6. 마상철, Vascular and Interventional Radiology, 대학서림, 2007.
  7. E. Vano, L. Arranz, and J. M. Sastre, "1998a. Dosimetric and radiation protection considerations based on some cases of patient skin injuries in interventional cardiology," Brit. J. Radiol, Vol.71, pp.510-516, 1998. https://doi.org/10.1259/bjr.71.845.9691896
  8. ICRP, Recommendations of the International Commission on Radiological Protection, ICRP Publication 26, Pergamon Press, Oxford, New york, 1997.
  9. 신성규, "골반촬영 시 내장된 필터를 이용한 환자 선량 감소", 한국콘텐츠학회논문지, 제12권, 제8호, pp.233-239, 2012.
  10. 유명규, "중재적 시술 시 Detector의 높이와 영상의 확대에 따른 환자의 흡수선량 변화에 관한 연구", 대한인터벤션영상기술학회지, 제13권, 제1호, pp.122-132, 2010.
  11. 이종석, "흉부 및 복부에서 AEC 적용에 따른 MDCT의 선량 감소 효과", 한국콘텐츠학회논문지, 제9권, 제3호, pp.225-231, 2009. https://doi.org/10.5392/JKCA.2009.9.3.225
  12. Roberto lezzi, "High concentration contrast media and radiation dose reduction in CT vascular imaging," Diagnostic Imaging Europe, Vol.5, pp.1-17, 2011.