Clinical gap changes after porcelain firing cycles of zirconia fixed dentures

  • Received : 2013.09.11
  • Accepted : 2014.03.05
  • Published : 2014.06.30

Abstract

PURPOSE. The aim of this study was to measure the changes on the marginal and internal adaptation of zirconia based anterior fixed partial dentures after the porcelain firing process. MATERIALS AND METHODS. A total of 34 anterior fixed partial dentures using LAVA CAD/CAM system (3M ESPE, Germany) were applied. Two silicone replicas were obtained: one is obtained before porcelain firing process (initial) and the other is obtained after porcelain firing process (final), followed by the examination under a binocular stereomicroscope. Kruskal Wallis and Wilcoxon Signed Ranks tests were used for the statistical analysis (P<.05). RESULTS. No statistically significant difference was found between initial and final marginal gap values (P>.05). At the internal gap measurements, final marginal area values ($59.54{\mu}m$) were significantly lower than the initial marginal area values ($68.68{\mu}m$)(P<.05). The highest and the lowest internal gap values were observed at the incisal/occlusal area and at the marginal area, respectively. In addition, lower internal gap values were obtained for canines than for central incisors, lateral incisors and premolars at the incisal area (P<.05). CONCLUSION. The firing cycles did not affect the marginal gap of Lava CAD/CAM system, but it is controversial for the internal gap.

Keywords

References

  1. Raigrodski AJ. Contemporary materials and technologies for all-ceramic fixed partial dentures: a review of the literature. J Prosthet Dent 2004;92:557-62. https://doi.org/10.1016/j.prosdent.2004.09.015
  2. Aboushelib MN, de Jager N, Kleverlaan CJ, Feilzer AJ. Microtensile bond strength of different components of core veneered all-ceramic restorations. Dent Mater 2005;21:984-91. https://doi.org/10.1016/j.dental.2005.03.013
  3. Yanagida H, Komoto K, Miyayama M. The chemistry of ceramics. Chichester; New York; Wiley; 1996.
  4. Piwowarczyk A, Ottl P, Lauer HC, Kuretzky T. A clinical report and overview of scientific studies and clinical procedures conducted on the 3M ESPE Lava All-Ceramic System. J Prosthodont 2005;14:39-45. https://doi.org/10.1111/j.1532-849X.2005.00003.x
  5. Holmes JR, Bayne SC, Holland GA, Sulik WD. Considerations in measurement of marginal fit. J Prosthet Dent 1989;62: 405-8. https://doi.org/10.1016/0022-3913(89)90170-4
  6. Sulaiman F, Chai J, Jameson LM, Wozniak WT. A comparison of the marginal fit of In-Ceram, IPS Empress, and Procera crowns. Int J Prosthodont 1997;10:478-84.
  7. Jacobs MS, Windeler AS. An investigation of dental luting cement solubility as a function of the marginal gap. J Prosthet Dent 1991;65:436-42. https://doi.org/10.1016/0022-3913(91)90239-S
  8. Goldman M, Laosonthorn P, White RR. Microleakage-full crowns and the dental pulp. J Endod 1992;18:473-5. https://doi.org/10.1016/S0099-2399(06)81345-2
  9. Jorgensen KD. Factors affecting the film thickness of zinc phosphate cements. Acta Odontol Scand 1960;18:479-90. https://doi.org/10.3109/00016356009043879
  10. Tinschert J, Natt G, Mautsch W, Spiekermann H, Anusavice KJ. Marginal fit of alumina-and zirconia-based fixed partial dentures produced by a CAD/CAM system. Oper Dent 2001;26:367-74.
  11. Boening KW, Wolf BH, Schmidt AE, Kastner K, Walter MH. Clinical fit of Procera AllCeram crowns. J Prosthet Dent 2000;84:419-24. https://doi.org/10.1067/mpr.2000.109125
  12. Reich S, Wichmann M, Nkenke E, Proeschel P. Clinical fit of all-ceramic three-unit fixed partial dentures, generated with three different CAD/CAM systems. Eur J Oral Sci 2005;113: 174-9. https://doi.org/10.1111/j.1600-0722.2004.00197.x
  13. Reich S, Kappe K, Teschner H, Schmitt J. Clinical fit of fourunit zirconia posterior fixed dental prostheses. Eur J Oral Sci 2008;116:579-84. https://doi.org/10.1111/j.1600-0722.2008.00580.x
  14. Gonzalo E, Suarez MJ, Serrano B, Lozano JF. Marginal fit of Zirconia posterior fixed partial dentures. Int J Prosthodont 2008;21:398-9.
  15. Vigolo P, Fonzi F. An in vitro evaluation of fit of zirconiumoxide- based ceramic four-unit fixed partial dentures, generated with three different CAD/CAM systems, before and after porcelain firing cycles and after glaze cycles. J Prosthodont 2008;17:621-6. https://doi.org/10.1111/j.1532-849X.2008.00366.x
  16. Gonzalo E, Suarez MJ, Serrano B, Lozano JF. A comparison of the marginal vertical discrepancies of zirconium and metal ceramic posterior fixed dental prostheses before and after cementation. J Prosthet Dent 2009;102:378-84. https://doi.org/10.1016/S0022-3913(09)60198-0
  17. Beuer F, Aggstaller H, Edelhoff D, Gernet W, Sorensen J. Marginal and internal fits of fixed dental prostheses zirconia retainers. Dent Mater 2009;25:94-102. https://doi.org/10.1016/j.dental.2008.04.018
  18. Wettstein F, Sailer I, Roos M, Hammerle CH. Clinical study of the internal gaps of zirconia and metal frameworks for fixed partial dentures. Eur J Oral Sci 2008;116:272-9. https://doi.org/10.1111/j.1600-0722.2008.00527.x
  19. Denissen H, Dozic A, van der Zel J, van Waas M. Marginal fit and short-term clinical performance of porcelain-veneered CICERO, CEREC, and Procera onlays. J Prosthet Dent 2000;84:506-13. https://doi.org/10.1067/mpr.2000.110258
  20. Bindl A, Mormann WH. Marginal and internal fit of all-ceramic CAD/CAM crown-copings on chamfer preparations. J Oral Rehabil 2005;32:441-7. https://doi.org/10.1111/j.1365-2842.2005.01446.x
  21. Reich S, Uhlen S, Gozdowski S, Lohbauer U. Measurement of cement thickness under lithium disilicate crowns using an impression material technique. Clin Oral Investig 2011;15: 521-6. https://doi.org/10.1007/s00784-010-0414-x
  22. Kokubo Y, Nagayama Y, Tsumita M, Ohkubo C, Fukushima S, Vult von Steyern P. Clinical marginal and internal gaps of In-Ceram crowns fabricated using the GN-I system. J Oral Rehabil 2005;32:753-8. https://doi.org/10.1111/j.1365-2842.2005.01506.x
  23. Castellani D, Baccetti T, Clauser C, Bernardini UD. Thermal distortion of different materials in crown construction. J Prosthet Dent 1994;72:360-6. https://doi.org/10.1016/0022-3913(94)90554-1
  24. Balkaya MC, Cinar A, Pamuk S. Influence of firing cycles on the margin distortion of 3 all-ceramic crown systems. J Prosthet Dent 2005;93:346-55. https://doi.org/10.1016/j.prosdent.2005.02.003
  25. Att W, Komine F, Gerds T, Strub JR. Marginal adaptation of three different zirconium dioxide three-unit fixed dental prostheses. J Prosthet Dent 2009;101:239-47. https://doi.org/10.1016/S0022-3913(09)60047-0
  26. Dittmer MP, Borchers L, Stiesch M, Kohorst P. Stresses and distortions within zirconia-fixed dental prostheses due to the veneering process. Acta Biomater 2009;5:3231-9. https://doi.org/10.1016/j.actbio.2009.04.025
  27. Kohorst P, Brinkmann H, Dittmer MP, Borchers L, Stiesch M. Influence of the veneering process on the marginal fit of zirconia fixed dental prostheses. J Oral Rehabil 2010;37:283-91. https://doi.org/10.1111/j.1365-2842.2009.02053.x
  28. Aboushelib MN, Feilzer AJ, de Jager N, Kleverlaan CJ. Prestresses in bilayered all-ceramic restorations. J Biomed Mater Res B Appl Biomater 2008;87:139-45.
  29. Pospiech PR, Rountree PR, Nothdurft FP. Clinical evaluation of zirconia-based all ceramic posterior bridges: two-year results. J Dent Res 2003;82:114.
  30. Raigrodski AJ, Chiche GJ, Potiket N, Hochstedler JL, Mohamed SE, Billiot S, Mercante DE. The efficacy of posterior three-unit zirconium-oxide-based ceramic fixed partial dental prostheses: a prospective clinical pilot study. J Prosthet Dent 2006;96:237-44. https://doi.org/10.1016/j.prosdent.2006.08.010
  31. Crisp RJ, Cowan AJ, Lamb J, Thompson O, Tulloch N, Burke FJ. A clinical evaluation of all-ceramic bridges placed in UK general dental practices: first-year results. Br Dent J 2008; 205:477-82. https://doi.org/10.1038/sj.bdj.2008.937

Cited by

  1. Verification of a computer-aided replica technique for evaluating prosthesis adaptation using statistical agreement analysis vol.9, pp.5, 2017, https://doi.org/10.4047/jap.2017.9.5.358