DOI QR코드

DOI QR Code

Vital Sign Sensor Based on Second Harmonic Frequency Drift of Oscillator

발진기의 2채배 고조파 주파수 천이를 이용한 생체신호 측정센서

  • Ku, Ki-Young (Technology Analysis Team, Defense Agency for Technology and Quality) ;
  • Hong, Yunseog (Radar and PGM Center, Hanwha Thales) ;
  • Lee, Hee-Jo (Department of Physics Education, Daegu University) ;
  • Yun, Gi-Ho (Department of Information and Communication Engineering, Sungkyul University) ;
  • Yook, Jong-Gwan (Department of Electric and Electronic Engineering, Yonsei University) ;
  • Kim, Kang-Wook (School of Electronics Engineering, College of IT Engineering, Kyoungpook National University)
  • 구기영 (국방기술품질원 기술분석팀) ;
  • 홍윤석 (한화탈레스 레이더체계그룹) ;
  • 이희조 (대구대학교 물리교육과) ;
  • 윤기호 (성결대학교 정보통신공학부) ;
  • 육종관 (연세대학교 전기전자공학과) ;
  • 김강욱 (경북대학교 전자공학과)
  • Received : 2015.11.11
  • Accepted : 2016.02.22
  • Published : 2016.03.31

Abstract

In this paper, a vital sign sensor based on impedance variation of resonator is proposed to detect the respiration and heartbeat signals within near-field range as a function of the separation distance between resonator and subject. The sensor consists of an oscillator with a built-in planar type patch resonator, a diplexer for only pass the second harmonic frequency, amplifier, SAW filter, and RF detector. The cardiac activity of a subject such as respiration and heartbeat causes the variation of the oscillation frequency corresponding impedance variation of the resonator within near-field range. The combination of the second harmonic oscillation frequency deviation and the superior skirt frequency of the SAW filter enables the proposed sensor to extend twice detection range. The experimental results reveal that the proposed sensor placed 40 mm away from a subject can reliably detect respiration and heartbeat signals.

본 논문에서는 공진기와 인체와의 거리에 따른 근접 전자기장 변화에 의한 공진기 임피던스의 변화를 토대로 사람의 호흡 및 심박신호를 검출할 수 있는 센서를 제안한다. 제안된 생체신호 측정센서는 패치형 공진기가 결합된 발진기, 발진주파수의 2채배 주파수만 통과시키기 위한 다이플렉서, 증폭기, SAW 필터 및 RF 검출기로 구성되어 있다. 호흡과 심박신호와 같은 인체의 주기적인 움직임은 근접 전자기장 영역 안에서 공진기의 임피던스 변화를 야기하며, 발진기의 주파수를 변화시킨다. 감도를 향상시키기 위해 발진주파수의 2채배 주파수 천이를 SAW 필터의 저지대역에 위치시킴으로써, 제안된 센서의 검출 거리를 2배로 확장시킬 수 있다. 제안된 센서의 측정결과, 최대 40 mm까지 호흡 및 심박신호가 안정적으로 측정되는 것을 확인하였다.

Keywords

References

  1. 이석희, 류근택, "유비쿼터스 헬스케어를 위한 무선 생체신호감시시스템 설계", 전자공학회논문지, 49(2), pp. 82-88, 2012년 6월.
  2. 장문석, 엄수홍, 김평수, 이흥혁, "u-헬스케어 시스템 에서 멀티 생체신호 측정 무선 단말기 설계 및 구현", 재활복지공학회 논문지, 7(2), pp. 27-34, 2013년 12월.
  3. A. Gruetzmann, S. Hansen, and J. Muller, "Novel dry electrodes for ECG monitoring", Journal of Physiol. Meas., vol. 28, no. 11, pp. 1375-1390, Oct. 2007. https://doi.org/10.1088/0967-3334/28/11/005
  4. Y. G. Lim, K. K. Kim, and K. S. Park, "ECG recording on a bed during sleep without direct skin-contact", IEEE Trans. Biomed. Eng., vol. 54, no. 4, pp. 718-725, Apr. 2007. https://doi.org/10.1109/TBME.2006.889194
  5. A. Rosen, M. A. Stuchly, and A. V. Vorst, "Application of RF/microwaves in medicine", IEEE Trans. Microw. Theory Techn., vol. 50, no. 3, pp. 963-974, Mar. 2002. https://doi.org/10.1109/22.989979
  6. L. Anitori, A. de Jong, and F. Nennie, "FMCW radar for life-sign detection", in Proc. IEEE Radar Conference, 2009, pp. 1-6.
  7. J. Han, J. -G. Kim, and S. Hong, "A compact Ka-band Doppler radar sensor for remote human vital signal detection", Journal of Electromagnetic Engineering and Science, vol. 12, no. 4, pp. 234-239, 2012. https://doi.org/10.5515/JKIEES.2012.12.4.234
  8. Y. -J. An, G. -H. Yun, and J.-G. Yook, "Wrist pulse detection system based on changes in the near-field reflection coefficient of a resonator", IEEE Microw. Wireless Compon. Letters, vol. 24, no. 10, pp. 719-721, Oct. 2014. https://doi.org/10.1109/LMWC.2014.2340584
  9. 안용준, 윤기호, 육종관, "휴대기기에서의 안테나 반사 신호를 이용한 심박 신호 검출", 한국통신학회논문지, 38(7), pp. 643-649, 2013년 7월.
  10. S. -G. Kim, G. -H. Yun, and J. -G. Yook, "Compact vital signal sensor using oscillation frequency deviation", IEEE Tran. Microw. Theory Tech., vol. 60, no. 2, pp. 393-400, Feb. 2012. https://doi.org/10.1109/TMTT.2011.2175403
  11. 김기윤, 김상규, 홍윤석, 육종관, "공진기의 임피던스 변화에 근거한 비접촉 생체 신호 센서", 한국통신학회논문지, 38(9), pp. 813-821, 2013년 9월.
  12. B. -H. Kim, Y. Hong, Y. -J. An, S. -G. Kim, G. -H. Yun, and J. -G. Yook, "Proximity coupled vital sign seonsor based on phase locked loop under injection", in Proc. IEEE Asia-Pacific Microw. Conf., 2013, pp. 536- 538.
  13. Y. Hong, S. -G. Kim, B. -H. Kim, S. -J. Ha, H. -J. Lee, G. -H. Yun, and J. -G. Yook, "Noncontact proximity vital sign sensor based on PLL for sensitivity enhancement", IEEE Trans. Biomedical Circuits and Systems, vol. 8, no. 4, pp. 584-593, Aug. 2014. https://doi.org/10.1109/TBCAS.2013.2280913
  14. C. Balanis, Antenna Theory Analysis and Design, 2nd ed. New York: Wiley, 1997.
  15. G. Gonzalez, Foundations of Oscillator Circuit Design, Artech House, Norwood, MA, 2007.
  16. 최재원, 이종민, 서철헌, "Microstrip square open loop 와 tunable negative resistance를 이용한 저위상 잡음 전압 제어 발진기", 한국전자파학회논문지, 17(12), pp. 1143- 1149, 2006년 12월.
  17. S. Gabriel, R. W. Lau, and C. Gabriel, "The dielectric properties of biolpgical tissues: II. Measurements in the frequency range 10 Hz to 20 GHz", Physics in Medicine and Biology, vol. 41, no. 11, pp. 2251, Apr. 1996. https://doi.org/10.1088/0031-9155/41/11/002
  18. J. -S. Hong, M. J. Lancaster, Microstrip Filters for RF/ Microwave Application, New York: Wiley, 2001, pp. 259-271.