The Study on Reduction of Image Distortion by using Single-Shot Turbo Spin Echo in Brain Stem Diffusion MRI

자기공명 확산강조영상검사 시 영상왜곡 감소에 관한 연구

  • Choi, Kwan-Woo (Department of Radiology, Asan Medical Center) ;
  • Lee, Ho-Beom (Department of Radiology, Asan Medical Center) ;
  • Na, Sa-Ra (Department of Radiology, Asan Medical Center) ;
  • Yoo, Beong-Gyu (Department of Radiotechnology, Wonkwang Health Science University) ;
  • Son, Soon-Yong (Department of Radiotechnology, Wonkwang Health Science University)
  • 최관우 (서울아산병원 영상의학과) ;
  • 이호범 (서울아산병원 영상의학과) ;
  • 나사라 (서울아산병원 영상의학과) ;
  • 유병규 (원광보건대학교 방사선과) ;
  • 손순룡 (원광보건대학교 방사선과)
  • Received : 2016.05.18
  • Accepted : 2016.06.30
  • Published : 2016.06.30

Abstract

Single-shot echo planar imaging(SS-EPI) is well established as high sensitivity for ischemic stroke. However, it is prone to susceptibility artifact in brain stem that diminish the image quality. single-shot turbo spin echo(SS-TSE) is a new DWI technique that can reduce susceptibility artifact. Thus, this research was conducted so as to reduce geometric distortion in brain stem by using single-shot turbo spin echo technique. Thirty patients without brain disease underwent diffusion MR on a 3T scanner with SS-EPI and SS-TSE. Obtained images with both sequences were analyzed for geometric distortion and error percentage as well. Image quality in terms of geometric distortion of SS-TSE were found to be significantly better than those for SS-EPI. And error percentage was considerably reduced for 2.4% of b0 image(from 11.1% to 8.7%), 1.2% of b1000 image(from 11.4% to 10.1%), respectively. In summary, diffusion MR using SS-TSE significantly reduce geometric distortion compared to SS-EPI in brain stem and may provide improved diagnostic performance.

본 연구는 DWI 검사 시 자화율 차이로 인해 뇌줄기에 발생하는 영상의 왜곡을 새로운 SS-TSE를 적용하여 감소시키고자 하였다. 연구방법은 2015년 7월부터 동년 9월까지 DWI 검사를 검사한 30명을 대상으로, 기존의 SS-EPI 기법과 새로운 SS-TSE를 적용하여 기법에 따른 뇌줄기의 왜곡과 오차율을 비교평가 하였다. 연구결과, 새로운 SS-TSE 적용 시 왜곡이 감소된 것을 볼 수 있으며, 오차율 또한 b0 영상은 2.4%(11.1%에서 8.7%), b1000 영상은 1.2%(11.4%에서 10.1%) 감소하는 것을 볼 수 있었다. 결론적으로, 뇌줄기의 DWI 검사 시 본 연구의 SS-TSE를 이용하면 기존 기법 사용 시 발생하는 영상의 왜곡을 감소시킬 수 있어 진단적 가치가 높은 영상을 획득할 수 있다.

Keywords

References

  1. D. Epstein, A. Mason, A. Manca, "The hospital costs of care for stroke in nine European countries", Health Economics, Vol. 17, No. 1, pp. 21-31, 2008. https://doi.org/10.1002/hec.1239
  2. Bogousslavsky, M. G. Van, F. Regli, "The Lausanne Stroke Registry: analysis of 1,000 consecutive patients with first stroke", Stroke, Vol. 19, No. 9, pp. 1083-1092, 1988. https://doi.org/10.1161/01.STR.19.9.1083
  3. K. M. Burger, S. Tuhrim, T. P. Naidich, "Brainstem vascular stroke anatomy", Neuroimaging Clinics of North America, Vol. 15, No. 2, pp. 297-324, 2005. https://doi.org/10.1016/j.nic.2005.05.005
  4. J. H. Lee, J. S. Lee, I. C. Im, "Evaluated the L-spine Magnetic Resonance Imaging for the Scanning Method of the Lateral Recumbent Position with to the Embarrassed Control of the Acute Low Back Pain", Journal of the Korean Society of Radiology, Vol. 8, No. 5, pp. 255-260, 2014. https://doi.org/10.7742/jksr.2014.8.5.255
  5. J. F. Burke, K. A. Kerber, T.J. Iwashyna, L. B. Morgenstem, "Wide variation and rising utilization of stroke magnetic resonance imaging", Annals of Neurology, Vol. 71, No. 2, pp. 179-185, 2012. https://doi.org/10.1002/ana.22698
  6. C. Z. Simonsen, M. H. Madsen, M. L. Schmitz, I. K. Mikkelsen, M. Fisher, G. Andersen, "Sensitivity of diffusion- and perfusion-weighted imaging for diagnosing acute ischemic stroke is 97.5%", Stroke, Vol. 45, No. 1, pp. 98-101, 2015.
  7. Y. L. Thian, W. Xie, D. A. Porter, A. B. Weileng, "Readout-segmented echo-planar imaging for diffusion-weighted imaging in the pelvis at 3T-A feasibility study", Academic Radiology, Vol. 21, No. 4, pp. 531-537, 2014. https://doi.org/10.1016/j.acra.2014.01.005
  8. S. J. Holdsworth, M. U. Antonucci, J. B. Andre, J. Rosenberg, M. Aksoy, M. Straka, N. J. Fischbein, R. Bammer, S. Skare, "Diffusion-weighted imaging with dual-echo echo-planar imaging for better sensitivity to acute stroke", American Journal of Neuroradiology, Vol. 35, No. 7, pp. 1293-1302, 2014. https://doi.org/10.3174/ajnr.A3921
  9. K. W. Yeom, S. J. Holdsworth, A. T. Van, R. M. Lober, R. Bammer, "Comparison of readout-segmented echo-planar imaging (EPI) and single-shot EPI in clinical application of diffusion-weighted imaging of the pediatric brain", American Journal of Roentgenology, Vol. 200, No. 5, pp. 437-443, 2013. https://doi.org/10.2214/AJR.12.8792
  10. R. Bammer, "Basic principles of diffusion-weighted imaging", European Journal of Radiology, Vol. 45, No. 3, pp. 169-184, 2003. https://doi.org/10.1016/S0720-048X(02)00303-0
  11. K. P. Pruessmann, M. Weiger, M. B. Scheidegger, P. Boesiger, "SENSE: sensitivity encoding for fast MRI", Magnetic Resonance in Medicine, Vol. 42, No. 5, pp. 952-962, 1999. https://doi.org/10.1002/(SICI)1522-2594(199911)42:5<952::AID-MRM16>3.0.CO;2-S
  12. K. O. Lovblad, P. M. Jakob, Q. Chen, A. E. Baird, G. Schlaug, S. Warach, R. R. Edelman, "Turbo spin-echo diffusion-weighted MR of ischemic stroke", American Journal of Neuroradiology, Vol. 19, No. 2, pp. 201-208, 1998.
  13. J. P. Vercruysse, B. Pilet, J. Michiels, R. Vertriest, M. Pouillon, T. Somers, E. Offeciers, "Single-shot, turbo spin-echo, diffusion-weighted imaging versus spin-echo-planar, diffusion-weighted imaging in the detection of acquired middle ear cholesteatoma", American Journal of Neuroradiology, Vol. 27, No. 7, pp. 1480-1482, 2006.
  14. J. S. Zhang, Y. Huan, "Multishot diffusion-weighted MR imaging features in acute trauma of spinal cord", European Journal of Radiology, Vol. 24, No. 3, pp. 685-692, 2014. https://doi.org/10.1007/s00330-013-3051-3
  15. T. Ishihara, K. Hirata, J. Kubo, K. Yamazaki, T. Sato, "Clinical application of multi-shot diffusion EPI inneurological disease", Rinsho Shinkeiqaku, Vol. 38, No. 5, pp. 453-456, 1998.
  16. F. Hennel, "Image-based reduction of artifacts in multishot echo-planar imaging", Journal of Magnetic Resonance, Vol. 134, No. 2, pp. 206-213, 1998. https://doi.org/10.1006/jmre.1998.1502
  17. M. H. Buonocore, L. Gao, "Ghost artifact reduction for echo planar imaging using image phase correction", Magnetic Resonance in Medicine, Vol. 38, No. 1, pp. 89-100, 1997. https://doi.org/10.1002/mrm.1910380114
  18. K. W. Choi, D. K. Seo, H. B. Lee, H. J. Goh, S. R. Na, "A study on image distortion improvement using silicon device in thyroid diffusion MRI images", Journal of the Korea Academia-Industrial cooperation Society, Vol. 15, No. 7, pp. 4380-4386. 2014. https://doi.org/10.5762/KAIS.2014.15.7.4380
  19. J. Pouwels, F. Rodjan, A. C. Moll, S. M. Imhof, D. L. Knol, E. Sanchez, J. A. Castelijns, "Single-shot turbo spin-echo diffusion-weighted imaging for retinoblastoma: initial experience", American Journal of Neuroradiology, Vol. 33, No. 1, pp. 110-118, 2012. https://doi.org/10.3174/ajnr.A2729

Cited by

  1. 악성종양의 감별진단을 위한 근골격의 확산강조영상 검사 시 최적의 여기횟수 vol.18, pp.8, 2016, https://doi.org/10.5392/jkca.2018.18.08.338
  2. 관자뼈의 확산강조영상검사 시 Single Shot Turbo Spin Echo 기법의 유용성 vol.44, pp.1, 2016, https://doi.org/10.17946/jrst.2021.44.1.25