Abstract

Background: A "difficult airway" can be simulated with an extrication collar, which restricts cervical motion and mouth opening. The purpose of this study is to compare the efficacy of the GlideScope and the McGrath in difficult airway simulation. Methods: Patients were randomized using computer-generated numbers and were placed into the GlideScope group or the McGrath group. The total intubation time was defined as the time measured from when the anesthesiologist picks up the device to the time at which three successive end-tidal $CO_2$ values are acquired after intubation. Results: There was no significant difference in total intubation time between the two groups ($73.0{\pm}25.3$ sec vs. $72.3{\pm}20.9$ sec, P = 0.92). The success rates of the first intubation attempt did not differ between the two groups (82.8% vs. 83.3%, P = 0.95). Conclusions: Our results suggest that there are no significant differences in the intubations with GlideScope and McGrath using vascular forceps and tube exchangers in difficult intubation scenarios.

Keywords

References

  1. Bathory I, Frascarolo P, Kern C, Schoettker P. Evaluation of the GlideScope for tracheal intubation in patients with cervical spine immobilisation by a semi-rigid collar. Anaesthesia 2009; 64: 1337-41. https://doi.org/10.1111/j.1365-2044.2009.06075.x
  2. Shippey B, Ray D, McKeown D. Use of the McGrath videolaryngoscope in the management of difficult and failed tracheal intubation. Br J Anaesth 2008; 100: 116-9. https://doi.org/10.1093/bja/aem303
  3. Theiler LG, Kleine-Brueggeney M, Kaiser D, Urwyler N, Luyet C, Vogt A, et al. Crossover comparison of the laryngeal mask supreme and the i-gel in simulated difficult airway scenario in anesthestized patients. Anesthesiology 2009; 111: 55-62. https://doi.org/10.1097/ALN.0b013e3181a4c6b9
  4. Phua DS, Mah CL, Wang CF. The The Shikani optical stylet as an alternative to the GlideScope$^{(R)}$ videolaryngoscope in simulated difficult intubations--a randomised controlled trial. Anaesthesia 2012: 67: 402-6. https://doi.org/10.1111/j.1365-2044.2011.07023.x
  5. Jeon WJ, Shim JH, Cho SY, Baek SJ. Stylet or forceps-guided tube exchanger to facilitate GlideScope intubation in simulated difficult intubations- a randomized controlled trial. Anaesthesia 2013; 68: 585-90. https://doi.org/10.1111/anae.12219
  6. Kheterpal S, Martin L, Shanks AM, Tremper KK. Prediction and outcomes of impossible mask ventilation: a review of 50,000 anesthestics. Anesthesiology 2009; 110: 891-7. https://doi.org/10.1097/ALN.0b013e31819b5b87
  7. Asai T, Koga K, Vaughan RS. Respiratory compications associated with tracheal intubation and extubation. Br J Anaesth 1998; 80: 767-75. https://doi.org/10.1093/bja/80.6.767
  8. Cattano D, Panicucci E, Paolicchi A, Forfori F, Giunta F, Hagberg C. Risk factors assessment of the difficult airway: an italiansurvey of 1956 patients. Anesth Analg 2004; 99: 1774-9.
  9. Combes X, Le Roux B, Suen P, Dumerat M, Motamed C, Sauvat S, et al. Unanticipated airway in anesthetized patients: prospective validation of a management algorithm. Anesthesiology 2004; 100: 1146-50. https://doi.org/10.1097/00000542-200405000-00016
  10. Savoldelli GL, Schiffer E, Abegg C, Baeriswyl V, Clergue F, Waeber JL. Comparison of the Glidescope, the McGrath, the Airtrap and the Macintosh laryngoscopes in simulated difficult airways. Anaesthesia 2008; 63: 1358-64. https://doi.org/10.1111/j.1365-2044.2008.05653.x
  11. Maassen R, Lee R, Hermans B, Marcus M, van Zundert A. A comparison of three videolaryngoscopes: the Macintosh laryngoscope blade reduces, but does not replace routine stylet use for intubation in morbidly obese patients. Anesth Analg 2009; 109: 1560-5. https://doi.org/10.1213/ANE.0b013e3181b7303a
  12. Jeon WJ, Kim KH, Teom JH, Bang MR, Hong JB, Cho SY. A comparison of the Glidescope$^{(R)}$ to the McGrath$^{(R)}$ videolaryngoscope in patients. Korean J Anesthesiol 2011; 61: 19-23. https://doi.org/10.4097/kjae.2011.61.1.19

Cited by

  1. Randomized comparison of the effectiveness of nasal intubation using a GlideScope video laryngoscope with Magill forceps versus vascular forceps in patients with a normal airway vol.64, pp.12, 2017, https://doi.org/10.1007/s12630-017-0971-4