DOI QR코드

DOI QR Code

Noise Reduction on Low Tube Voltage CT Images

저관전압 CT영상에서 발생되는 노이즈 제거

  • Choi, Seokyoon (Department of Radiological Science, Catholic University of Pusan)
  • 최석윤 (부산가톨릭대학교 보건과학대학 방사선학과)
  • Received : 2017.02.15
  • Accepted : 2017.02.28
  • Published : 2017.02.28

Abstract

To reduce the exposure dose in head CT, the use of low tube voltage is required. However, increasing noise may cause errors in the second data processing. In this study, we proposed a method to reduce noise by using low tube voltage. Experimental results show that the noise level is high at 100kVp and lowest at 140 kVp. The dose was lower at 100 kVp and higher at 140 kVp. As a result of applying the wavelet according to the threshold value, the noise value in the wavelet Th30 decreased to 4.51. Using the parameter condition(100 kVp, rotation time 0.5 sec, dose: 40.64 mGy) and the wavelet Th 30, the dose reduction of 65.3% was possible. We believe that applying the proposed method to head CT images will help to patient safety and interpret accurate information.

두부 CT에서 피폭선량을 줄이기 위해서 저관전압 사용이 요구된다. 그러나 노이즈가 증가하는 현상이 발생해서 2차 데이터 가공 시 오류를 초래할 수 있다. 본 연구에서는 저관전압을 사용하고 노이즈를 줄일 수 있는 방법을 제시하였다. 실험 결과 100kVp 에서 노이즈 값이 높게 나타났고, 140kVp에서 가장 낮게 나타났다. 100kVp에서 선량이 낮게 나타났고, 140kVp에서 선량이 높게 기록되었다. 역치값에 따른 웨이블릿의 적용 결과 웨이블릿 Th30에서 노이즈값은 4.51로 감소하였다.(50% 감소) 100kVp, 회전시간 0.5 sec (선량: 40.64 mGy)파라미터 조건과 웨이블릿 Th 30을 사용하면 65.3%의 피폭선량 감소가 가능하였다. 연구에서 제시한 방법을 두부CT영상에 적용한다면 환자의 안전과 정확한 정보를 해석하는데 도움을 줄 것이라고 판단한다.

Keywords

References

  1. M. Y. Kim, H. S. Kim, "The Evaluation of Radiation Dose by Exposure Method in Digital Magnification Mammography", JRST, Vol. 35, No. 4, pp. 293-298, 2012.
  2. S. Obenauer, K. P. Hermann and E Grabbe, "Dose reduction in full-field digital mammography: an anthropomorphic breast phantom study", Br J Radiol, Vol.76, No. 907, pp. 478-82, 2003 https://doi.org/10.1259/bjr/67597156
  3. I. J. Lee, "Relationship of Compressed Breast Thickness and Average Glandular Dose According to Focus /Filter", JRST, Vol. 32, No. 3, pp. 261-270, 2009
  4. J. Cho, H. Kang, H. Kim, S. Kim. Mulimedia Signal processing, SciTech media, 2006.
  5. R.Gonzale, R. E.Woods . Image processing, Green, Seoul, 2003.
  6. H. S. Cho, H. Lee, "Choise of Wavelet-Thresholds for Denoising images", The KIPS Transactionsty, Vol. 8, No. 2, pp. 693-698, 2001.
  7. M. B. Williams, P. Raghunathan, M. J. More, J. A. Seibert, A. Kwa, J. Y. Lo, E. Samei, N. T.Ranger, and L. L. Fajardo, L.L. "Optimization of exposure parameters in full field digital mammography. Med Phys", Vol. 35, pp. 2414-2423, 2008 https://doi.org/10.1118/1.2912177
  8. D. L. Donoho and J. M. Johnstone, "Ideal spatial adaptation via wavelet Shrinkage", Biometrica, Vol. 81, pp. 425-455, 1994. https://doi.org/10.1093/biomet/81.3.425
  9. S. Choi, S. J. Ko, S. S. Kang, "Denoising of Digital Mammography Images Using Wavelet Transform", J KSR, Vol. 7, No. 3, pp. 181-189, 2013.

Cited by

  1. 두부 CT 선량감소를 위한 총변량 최적화의 적용 vol.12, pp.6, 2017, https://doi.org/10.7742/jksr.2018.12.6.707
  2. 두부 CT검사에서의 노이즈 및 화질분석 vol.42, pp.4, 2017, https://doi.org/10.17946/jrst.2019.42.4.279
  3. 두부 CT의 노출 파라메타에 따른 화질과 선량의 변화 vol.13, pp.5, 2019, https://doi.org/10.7742/jksr.2019.13.5.705
  4. CT 영상의 화질개선을 위한 이중트리복합웨이블릿의 적용 vol.13, pp.7, 2017, https://doi.org/10.7742/jksr.2019.13.7.941
  5. 나선형영상획득에서 Pitch에 따른 CT 감약계수와 잡음의 변화 vol.14, pp.7, 2020, https://doi.org/10.7742/jksr.2020.14.7.981