DOI QR코드

DOI QR Code

Noise Changes in Image Plate

시간변화에 따른 영상판의 노이즈 변화

  • Choi, Seokyoon (Department of Radiological Science, Catholic University of Pusan) ;
  • Im, In-Chul (Department of Radiological Science, Dongeui University)
  • 최석윤 (부산가톨릭대학교 방사선학과) ;
  • 임인철 (동의대학교 방사선학과)
  • Received : 2017.06.12
  • Accepted : 2018.04.30
  • Published : 2018.04.30

Abstract

It is possible to do not use the Image Plate for a long time in the clinic. for the acquisition of reliable images the decommissioning process is very important. In this study, we would like to look at complex noises that may occur during Image Plate use depending on the frequency and time of day. first, we got the image from computer radiography. second, We calculate the noise changes using Peak Signal to Noise Ratio(PSNR) and Noise Power Spectrum(NPS). finally, we suggest that remove the noise from the IP for more than 4 hours. and It turned out to be better to remove the noise more than once. this study will be a quantitative reference for Image Plate management and use in the clinical trial.

컴퓨터방사선촬영 시스템이 도입되면서 영상판은 분명 우리에게 많은 정보, 정밀한 디지털이미지 제공, 재촬영률 감소 등 상당한 이점으로 다가왔다. 컴퓨터방사선촬영 시스템의 영상판을 이용한 의료 영상 획득시 좋은 영상을 얻기 위해서는 영상판의 보관 장소 및 발광시간의 응답 특성과 잠상퇴행 현상을 인지함과 동시에 효과적인 정도관리와 화질 향상을 위해 체계적인 관리와 노력을 해야 할 것으로 판단한다. 본 연구에서는 컴퓨터방사선촬영 시스템의 영상판을 이용한 영상획득 과정에서 발생할 수 있는 자연 환경에 의한 잡음에 대하여 알아보았다. 실험 결과 시간이 지남에 따라 영상판에 잡음이 생성되며 영상의 신뢰성이 떨어지게 됨을 확인하였다. 영상판을 사용할 경우에는 4시간 이상 보관된 영상판에 대해서는 잠상소거를 반드시 실시해서 사용하고, 2회 이상 잠상소거 하는 것이 가장 좋은 방법일 것이라고 판단한다. 본 연구는 임상에서 영상판 관리 및 사용 시 정량적인 참고자료가 될 것이다.

Keywords

References

  1. B. S. Kang, "Assessment of dose effects on image quality at chest computed radiography," The Korean Society of Radiology, Vol. 5, No 6, pp. 421-426, 2011. https://doi.org/10.7742/jksr.2011.5.6.421
  2. K. R. Dong, J. G. Choi, S. I. Hong, "Cause and Solution of an Artifact Generation by Parameter in Computed Radiography System," Journal of the Korean contents Association, Vol. 9, No 5, pp. 421-426, 2009.
  3. M. S. Yu, J. S. Lee, I. C. Im, "Evaluation of Approximate Exposure to Low-dose Ionizing Radiation from Medical Images using a Computed Radiography (CR) System," Journal of the Korean Society of radiology, Vol. 6, No 6, pp. 455-464, 2012. https://doi.org/10.7742/jksr.2012.6.6.455
  4. B. Y. Lim, H. S. Park, J. H. Kim, K. H. Park, H. J. Kim, "Evaluation of Unexposed Images after Erasure of Image Plate from CR System,"Journal of the Korean Physical Society, Vol. 20, No 4, pp. 199-207, 2009.
  5. S. B. Oh, S. M. Kwan, K. S. Chon, "Reduction of noise distribution according to the number of latent image erase of Image Plate," The Korean Society of Radiology, 2012 spring conference summary, pp. 37-39, 2012.
  6. J. W. Min, J. M. Kim, H. W. Jeong, "Mixed Noise Reduction Filters for CR Images," Journal of radiological science and technology, Vol. 30, No. 1, pp. 1-6, 2007.
  7. B. J. Kwak, T. J. Ji, "Evaluating Picture Quality of Image Plates in Digital CR Systems," Journal of Radiation Protection, Vol. 36, No. 4, pp. 220-221, 2011.