Comparison of Dose Measurement of Glass Dose Meter, Semiconductor Dose Meter, and Area Dose Meter in Diagnostic X-ray Energy

진단영역 X선 에너지에서 유리선량계, 반도체선량계, 면적선량계의 선량 실측 비교

  • Son, Jin-Hyun (Department of Radiological technology, Shingu University)
  • Received : 2019.11.28
  • Accepted : 2019.12.26
  • Published : 2019.12.31

Abstract

This paper obtained and compared these dose values by setting and comparing the X-ray imaging conditions (tube voltage 60 kVp, 70 kVp, 80 kVp, tube current 10 mAs, 16 mAs and X-ray field size are 10 × 10 cm, 15 × 15 cm). Each dose value was measure 10 times and represented as an average value. The purpose of this experiment is to serve as a reference for the X-ray exposure of diagnostic areas according to the type of dosimeter and to help with another dose measurement. The results of the experiment showed very little difference between the glass dosimeter(GD) and semiconductor dosimeter values due to changes in tube voltage of 60, 70, 80 kVp, regardless of field sized, but for dose area product(DAP), the difference in dose value was significant according to field size.

Keywords

References

  1. Kim JM, Do GH, Kim GP, Hwang JY, Choi HJ, Kim SG. A Study on building the management system and methods of patient radiation exposure dose. National Evidence-based Healthcare Collaborating Agency; 2014.
  2. KFDA. 2012 Guideline on patient dose suggestion of general radiology. Seoul, Korea, 2012.
  3. KFDA. Measuremant Guideline of patient dose. Seoul, Korea, 2007.
  4. Son JH, Min JH, Kim HS, Lyu KY, Lim HS, Kim JM, et al. Experimental study with respect to dose characteristic of glass dosimeter for low-energy by using internal detector of piranha 657. Journal of Radiological Science and Technology. 2012;35(2): 119-24.
  5. Kim JK, Kim JK. Evaluation on organ dose and image quality of lumbar spine radiography using glass dosimeter. Journal of Radiological Science and Technology. 2016;39(1):1-11. https://doi.org/10.17946/JRST.2016.39.1.01
  6. Hwang JH, Lee KB. A Study on the usefulness of glass dosimeter in the evaluation of absorbed dose by comparing the doses of multi-purpose dosimeter and glass dosimeter using kerma with PCXMC 2.0 in DR(Digital Radiography). The Korea Contents Society. 2017;17(9):292-9.
  7. Son JH, Min JH, Kim KW, Son SY, Lim HS. Experimental study on build up characteristic of glass dosimeter of preheat and Non-preheat in low energy according to delay time. Journal of the Korea Academia-Industrial cooperation Society. 2013; 14(7):3412-8. https://doi.org/10.5762/KAIS.2013.14.7.3412
  8. Choi JH, Kang GJ, Chang SG. Comparison on the dosimetry of TLD and PLD by dose area product. The Korea Contents Society. 2012;3:244-50. https://doi.org/10.5392/JKCA.2012.12.03.244
  9. Jung JE, Won DY, Jung HM, Kweon DC. Calibration examination of dose area product meters using X-ray. Journal of the Korean Society of Radiology. 2017;11(1):37-42. https://doi.org/10.7742/jksr.2017.11.1.37
  10. Kwak DH, Lee SH, Lee SH. Development of DAP(Dose Area Product) for radiation evaluation of medical and industrial X-ray generator. Institute of Korean Electrical and Electronics Engineers. 2018;22(2):495-8.
  11. Tung CJ, Tsai HY. Evaluations of Gonad and Fetal Doses for Diagnostic Radiology. Council Republic Of China Part B. 1993;23(3):107-13.
  12. RTI Electronics Inc. Piranha Reference Manual (Version 4.3A (10), 21), 2012.
  13. Asahi Techno Glass Corporation. Explanation material of RPL glass dosimeter: Small element system. Tokyo, Japan; 2000.
  14. Han JB, Choi NG, Sung HJ. Comparative study of radiation exposure using entrance skin dose calculation technique in diagnostic X-ray radiography. The Korea Contents Society. 2011;(12):357-63.
  15. Yun YW, Je JY. Evaluation of area dose product and image density according to the variable tube current. Journal of the Korean Society of Radiology. 2018:12(67):645-50.
  16. Lee GM, Kang BS, Park HH, Lee GY. An analysis of the area dose meter calibration and error rate in an IVR device. Journal of Korean Society of Cardio-Vascular Interventional Technology. 2012;15(1):128-34.
  17. Kim KT, Choi IH, Jung BJ, Kang SS, Noh SC, Heo YJ, et al. Feasibility study of DAP-based radiation dose management system in medical imaging system. The Institute of Electronics and Information Engineers. 2015;6:1157-9.
  18. Son JH, Kim SH, Mun HJ, Kim LK, Son IH, Kim YJ, et al. A study on characteristic of glass dosimeter according to graded change of tube current. Journal of Radiological Science and Technology. 2014;37(2):135-41.
  19. Kim KW, Son JH. Comparisons and measurements the dose value using the semiconductor dosimeter and dose area product dosimeter in skull, chest and abdomen. Journal of Radiological Science and Technology. 2015;38(2):101-6. https://doi.org/10.17946/JRST.2015.38.2.03