DOI QR코드

DOI QR Code

A Study on Performance Evaluation for Electrocardiography Signal Measurement Electrode based on Conductive Fabric

전도성 섬유 기반 심전도 전극의 성능 평가에 관한 연구

  • Kang, Bo Kyu (Graduate School of Biomedical Engineering, Yonsei University) ;
  • Yoo, Sun Kook (Department of Medical Engineering, College of Medicine, Yonsei University)
  • 강보규 (연세대학교 일반대학원 생체공학협동과정) ;
  • 유선국 (연세대학교 의과대학 의학공학교실)
  • Received : 2012.11.09
  • Published : 2013.02.25

Abstract

Recently, as we move toward a society with an increasingly aged population, wearable U-health devices in various shapes with smart wear have been developed in order to conveniently measure health variables without using hands in daily life or at home. However, the problem is that only supply of the wearable U-health devices is focused and its applicable devices are studied and developed, which has resulted in lack of awareness of importance of performance evaluation. In this study, two electrodes were fabricated using conductive fabric which can be used as electrode if attached to wearable U-health devices or smart wear in order to measure ECG signal. Two electrodes those were fabricated using conductive fabric were compared the correlation, impedance and CMRR with patch typed Ag-AgCl electrode-normally used for measurement of ECG signal, so that the study would find out if the fabricated electrode can be used with the wearable U-health devices by testing and evaluating performances.

인구의 고령화 사회에 접어든 최근에는 스마트 의류와 접목된 다양한 형태의 착용형 유헬스 디바이스가 개발되어 일상생활과 가정에서의 생활에 양손을 자유로이 하여 불편함 없이 건강변수 측정이 가능하게 되었다. 하지만 착용형 유헬스 디바이스의 보급과 더불어 이를 응용한 디바이스에 대한 연구 및 개발만이 우선되고 있는 현재, 이에 대한 성능 평가의 중요성에 대한 인식은 미흡한 실정이다. 본 논문에서는 이러한 착용형 유헬스 디바이스와 스마트 의류에 접목하여 전극으로 사용될 수 있는 전도성 섬유를 이용한 전극을 구성하여 심전도(ECG) 신호 측정을 위해 일반적으로 사용되는 패치타입의 Ag-AgCl 전극과 임피던스(Impedance), 심전도(ECG) 신호의 상관관계 비교와 CMRR 측정을 통하여 착용형 유헬스 디바이스에의 사용 가능성을 밝히기 위해 시험을 통한 성능차이를 평가 하였다.

Keywords

References

  1. A Searle and L Kirkup, "A direct comparison of wet, dry and insulting bioelectric recording electrodes," Physiol. Meas. 21, pp. 271-283, 2000. https://doi.org/10.1088/0967-3334/21/2/307
  2. Lun-De Liao, I-Jan Wang, Sheng-Fu Chen, jyh-Yeong Chang and Chin Teng Lin, "Design, Fabrication and Experimental Validation of a Novel Dry-Contact Sensor for Measuring Electroencephalography Signals without Skin Preparation," Sensors, 11, pp. 5819-5834, 2011. https://doi.org/10.3390/s110605819
  3. Junchae Kim, Sun Ha Jee and Sun Kook Yoo, "Fabrication and Evaluation of Sensor for Measuring Pulse Wave Velocity using Piezo Film and Conductive Textile," Journal of Sensor Science and Technology, Vol. 21, No. 2, pp. 135-143, 2012. https://doi.org/10.5369/JSST.2012.21.2.135
  4. Bo Kyu Kang, In Ho Hwang, Sun K. Yoo " A Study on Performance Evaluation for ECG Device Using Wearable u-Health Smart Cloth and Belt," Bio-Medical System Conference and Ohsong Bio-Forum(BIT), pp. 148-151, Feb 17-19, 2011.
  5. 이석희, 류근택, "유비쿼터스 헬스케어를 위한 무선 생체신호 감시 시스템 설계," 대한전자공학회 논문지, 제49권 IE편, 제2호, 82-88쪽, 2012.6
  6. 장용원, 신승철, 전영주, 김승환, "전도성 섬유를 이용한 운동중 생체 신호 측정 및 분석," 대한 전자공학회 추계 학술대회 논문집ΙΙ, 800-803쪽, 2006.
  7. 이학성, "임피던스 관계 행렬을 이용한 고집적 센서 어레이의 임피던스 분석에 관한 연구," 대한 전자공학회, 48권, SC편, 제3호, 1-5쪽, 2011년 5월.
  8. Enzo Pasquale Scilingo, Angelo Gemignani, Rita Paradiso, Nicola Taccini, Brunello Ghelarducci and Danilo De Rossi, "Performance Evaluation of Sensing Fabrics for monitoring Physiological and Biomechanical Variables," IEEE Transaction on Information technology in Biomedicine, Vol. 9, no. 3, pp. 345-352, September 2005. https://doi.org/10.1109/TITB.2005.854506
  9. Long Yan and Hoi-Jun Yoo, "A Low-Power Portable ECG Touch with Two Dry Metal Contact Electrodes," Journal of Semiconductor Technology and Science, Vol.10 No.4, pp. 300-308, 2010. 12. https://doi.org/10.5573/JSTS.2010.10.4.300
  10. 민철홍, 김태선, "심전도 신호의 리드III 파형을 이용한 바이오 인식." 전자공학회 논문지. 제 48권, 6 호, 43-50쪽, 2011년 11월.
  11. Merja M. Puurtinen, Satu M. Komulainen, Pasi K. Kauppinen, Jaakko A. Malmivuo and Jari A. K. Hyttinen, "Measurement of noise and impedance of dry and wet textile electrodes, and textile electrodes with hydrogel," Proceeding of the 28th IEEE, EMBS Annual International Conference, pp. 6012-6015, New York City, USA, Aug 30-Sept 3, 2006
  12. Israel Zepeda-Carapia, Agustin Marquez- Espinoza and Carlos Alvarado-Serrano, "Measurement of Skin-Electrode Impedance for a 12-Lead Electrocardiogram," 2nd International Conference on Electrical and Electronics Engineering(ICEEE) and XI Conference on Electrical Engineering(CIE 2005), pp.193-195, Mexico City, Mexico, September 7-9, 2005.
  13. Sun K. Yoo " Relative Measurement of Differential Electrode Impedance for Contact Monitoring in a Biopotential Amplifier," International Journal of Control, Automation, and System, Vol. 5, no. 5, pp601-605, October 2007.
  14. Yu Mike Chi, Tzyy-Ping Jung, and Gert Cauwenbrghs, "Dry-Contact and Noncontact Biopotential Electrodes: Methodological Review," IEEE Reviews in Biomedical Engineering, Vol. 3, 2010.
  15. Fabrie Axisa, Pierre Michael Schmitt, Claudine Gehin, Georges Delhomme, Eric McAdams, and Andre Dittmar, "Flexible Technologies and Smart Clothing for Citizen Medicine, Home Healthcare, and Disease Prevention," IEEE Transaction on Information Technology in Biomedicine, Vol. 9, No. 3, September 2005.
  16. Kang-Hwi Lee, Jeong-Whan Lee, Yong-Jae Lee, Kyeong-Seop Kim, Heui-Koung Yang, Kun-Su Shin, Myoung-Ho Lee, "Implementation of Wearable Heart Activity Monitoring System Having Modified Bipolar Electrode and Correlation Analysis with Clinical Electrocardiography(ECG)," Trnas. KIEE, Vol. 57, No. 6, pp. 1102-1108, Jun, 2008.
  17. Deokwon Ko, Chany Lee, Eun-Joong Lee, Sang-Hoon Lee and Ki-Young Jung, "A Dry and Flexible Electrode for continuous-EEG Mornitoring Using Silver Balls Based Polydimethylsiloxane(PDMS)," Biomed Eng Lett, Vol. 2, pp. 18-23, 2012. https://doi.org/10.1007/s13534-012-0049-8

Cited by

  1. Design of a Readout Circuit of Pulse Rate and Pulse Waveform for a U-Health System Using a Dual-Mode ADC vol.50, pp.9, 2013, https://doi.org/10.5573/ieek.2013.50.9.068