The Evaluation of Space Dose Distribution for Digital Mammography Equipment

디지털 유방촬영장비에 관한 공간선량율 평가

  • Jung, Hongmoon (Department of Radiological Science, Daegu Health College) ;
  • Jung, Jaeeun (Department of Radiological Science, Daegu Health College) ;
  • Hyun, Hyejin (Department of Medical laboratory Science, Daegu Health College) ;
  • Won, Doyeon (Department of Radiological Science, Daegu Health College)
  • 정홍문 (대구보건대학교 방사선과) ;
  • 정재은 (대구보건대학교 방사선과) ;
  • 현혜진 (대구보건대학교 임상병리과) ;
  • 원도연 (대구보건대학교 방사선과)
  • Received : 2014.11.24
  • Accepted : 2015.01.25
  • Published : 2015.01.30

Abstract

Mammography equipment is an essential detector for making an early diagnosis of female's breast lesion. Recently, in most hospitals, a digital mammography detector is used due to the wide and consistent supply of digital mammography equipment. However, the average effective radiation is increasing due to the indiscreet use of CR or DR mammography. The purpose of this study is to recognize the possible indirect radiation damage, which can be occurred due to an excessive effective exposure of radiation, by evaluating spacial radiation rate of the digital mammography detector used for female patient. Consequently, the high mount of spacial radiation showed digital mammography equipment on the horizontal direction. Considering the result, digital mammography equipment should be installed by avoiding along the horizontal direction.

유방촬영장비는 여성의 조기 유방의 병변을 진단하기 위하여 필수적으로 사용되는 장비이다. 계속적인 디지털 유방촬영장치가 보급되어짐에 따라 현재에는 대부분의 의료기관에서 디지털 유방촬영장치를 사용하고 있다. 하지만 이러한 디지털장비인 CR 또는 DR 유방촬영술의 무분별한 사용으로 여성 환자의 평균유효선량이 증가되고 있다. 이번 연구는 여성환자에게 사용되는 디지털 유방촬영장비의 공간선량율을 평가하여 방사선유효선량 과다 피폭으로 인하여 발생할 수 있는 방사선 2차 피해을 인식시키고자 하였으며, 더 나아가 디지털 유방촬영장치에서 발생되는 공간선량율 측정함으로써 환자에게 직접 촬영을 하는 방사선사 안전의식을 인식시키고자 한다. 결론적으로 디지털 유방촬영장비의 공간선량은 촬영장비를 기준으로 수평으로 높은 수치를 나타내었다. 이러한 결과를 고려한다면 디지털 유방촬영장비의 설치시에 수평 방향을 피하여 방사선사의 조작위치를 변경하거나 디지털 유방촬영 장비로부터 수평으로 되는 부위에 납유리차폐의 보완이 이루어져야 할 것이다.

Keywords

References

  1. Kim GS , Gang SH ,Lee, SJ , "Early Results of Subcutaneous Mastectomy with Immediate Breast Reconstruction as a Treatment for Early Breast Cancer", korean surgical society, Vol.68 No.4, 264-270.2005.
  2. Jeon MS, Moon SM, Lee HJ , Lee EH, Song Y,S et al, "Arm Morbidity after Breast Cancer Treatments and Analysis of Related Factors" , The Journal of the Korean soceity for therapeutic radiology and oncology , vol.23 ,no.1, pp.32-42.2005.
  3. Kim HC, Jo PG, Kim SS, Choi JH, Kim YH, "A Survey on Radiation Exposure of Patient in Mammography", JJournal of radiological science and technology , Vol. 27 no.4, pp.55-60, 2004.
  4. Kim TH, " Dosimetric Evaluation of Average Glandular Absorption Radiation Dose in Mammography ", Journal of the Korean radiological society , Vol. 35 no.6, pp.999-1003,1996. https://doi.org/10.3348/jkrs.1996.35.6.999
  5. Choi JH, Jeon MJ, Kim YI, Choi JW, "A Study on Radiation Dose in Mammography", Journal of Korean Society of radiological technology , Vol.4 no.1, pp.31-36,1981,
  6. Lee JS, " Dose Evaluation Using Mathematical Simulation of Radiation Exposure Body in Mammography System ", Journal of the Korean radiological society , Vol. 8 no.4,pp.155-161,2014. https://doi.org/10.7742/jksr.2014.8.4.155
  7. Jeong HR, Hwang Sy, Ha BC . Kim BH, Lee DW, et al, " The Average Glandular Dose in Mammography and Quality Control of the Equipment Status", Journal of the Korean radiological society, Vol. 5 no.3, pp.111-120,2011. https://doi.org/10.7742/jksr.2011.5.3.111
  8. Jo WI, Kim YG, Lee GD, "Change of Dose Exposure and Improvement of Image Quality by Additional Filtration in Mammography " ,Journal of radiation protection, Vol. 38 no.2,pp.78-90. 2013, https://doi.org/10.14407/jrp.2013.38.2.078

Cited by

  1. 유방X선촬영 시 피폭선량 감소를 위한 유방촬영용 차폐복의 유용성에 관한 연구 vol.42, pp.2, 2019, https://doi.org/10.17946/jrst.2019.42.2.99