DOI QR코드

DOI QR Code

Analysis of Magnetic Field Variation for Pulse Wave by Using Finite Element Method

유한 요소법을 이용한 맥진 파형의 자기장 변화 분석

  • Park, Tae-Young (Department of Oriental Biomedical Engineering, Sangji University) ;
  • Lee, Sang-Suk (Department of Oriental Biomedical Engineering, Sangji University)
  • 박태영 (상지대학교 보건과학대학 한방의료공학과) ;
  • 이상석 (상지대학교 보건과학대학 한방의료공학과)
  • Received : 2014.02.01
  • Accepted : 2014.04.29
  • Published : 2014.06.30

Abstract

The magnetic field variation by the permanent magnet fluctuation positioned on a "Chwan" of wrist according to the movement of radial artery was generated. The clip-type pulsimeter equipped with a Hall device sensing magnetic field pulse movement analyzed the characteristics of pulse wave as output signals. The magnetic field curve and pulse waveform simulated by the finite element method were compared and analyzed with each other. Also, the variation of magnetic field distribution one permanent magnet investigated by the fabrication of clip-type pulsimeter simulator. This result suggests that the clip-type pulsimeter can be used the reproducible and efficacious oriental diagnostic medical instrument.

요골동맥의 움직임에 따라 손목의 "관" 위치에 놓인 영구자석이 요동함으로 자기장의 변화를 생성하였다. 맥동에 의존하는 자기장을 감지하는 홀소자가 구비된 집게형맥진기의 출력신호인 맥진파형 특성을 분석하였다. 유한요소법으로 전산모사한 자기장 파형과 실제 맥진 파형의 시간 특징점들은 거의 일치하였다. 한 개의 원통 소형 영구자석의 자기장 분포의 변화를 감지하는 홀소자 집게형맥진기 시뮬레이터를 제작하여 재현성과 유효성 있는 한방진단기기로 활용하는 기틀을 마련하였다.

Keywords

References

  1. M. L. Chan, G. Jaramillo, K. R. Hristova, and D. A. Horsley, Biosens Bioelectron. 26, 2060 (2011). https://doi.org/10.1016/j.bios.2010.09.002
  2. P. Khajidmaa and S. S. Lee, J. Kor. Magn. Soc. 23, 193 (2013). https://doi.org/10.4283/JKMS.2013.23.6.193
  3. N. L. Anderson, A. Jackson, D. Smith, D. Hardie, C. Borchers, and T. W. Pearson, Mol. Cell. Proteomics 8, 995 (2009). https://doi.org/10.1074/mcp.M800446-MCP200
  4. S. M. Choi, Performance evaluation of medical devices in oriental medicine, Seoul: Ministry of Food and Drug Safety (2004).
  5. M. S. Han and I. H. Sung, J. Business Education 23, 385 (2009).
  6. Y. H. Bak, J. W. Kwon, D. S. Huang, and H. K. Shin, J. Korean Oriental Med. 32, 79 (2011).
  7. J. S. Moon, J. Y. Kim, Y. H. Ryu, N. K. Song, H. J. Kim, and J. R. Lee, J. Korean Oriental Med. 11, 119 (2005).
  8. Korea Institute of S&T Evaluation and Planning, Trend Brief Report, Technology and Policy Trend of Korean Medicine Science (2010. 8).
  9. K. J. Lee, J. Kor. Association of Applied Economics 14, 149 (2012).
  10. S. S. Lee, J. G. Choi, I. H. Son, K. H. Kim, D. H. Nam, Y. S. Hong, W. B. Lee, D. G. Hwang, and J. R. Rhee, J. Magn. 16. 449 (2011). https://doi.org/10.4283/JMAG.2011.16.4.449
  11. S. S. Lee, I. H. Son, J. G. Choi, D. H. Nam, Y. S. Hong, and W. B. Lee, J. Korean Phys. Soc. 58, 349 (2011). https://doi.org/10.3938/jkps.58.349
  12. Cobham technical services, Opera-3d User Guide, Version 15 (2011).
  13. B. W. Schaaf and P. H. Abbrecht, J. Biomechanics 5, 345 (1972). https://doi.org/10.1016/0021-9290(72)90064-4
  14. M. F. P. O'Rourke, R. P. Kelly, and A. P. Avolio, Arterial Pulse, 1st ed., Lea & Febiger, Philadelphia, USA (1992).
  15. G. D. Yoo, S. K. Hwang, and S. S. Lee, J. Korean Magn. Soc. 22, 27 (2012). https://doi.org/10.4283/JKMS.2012.22.1.027
  16. D. H. Lee, Y. S. Hong, and S. S. Lee, J. Korean Magn. Soc. 23. 1 (2013). https://doi.org/10.4283/JKMS.2013.23.1.001
  17. K. H. Kim and S. S. Lee, New Physics 63, 1135 (2013).
  18. I. H. Son and S. S. Lee, J. Korean Magn. Soc. 21, 104 (2011). https://doi.org/10.4283/JKMS.2011.21.3.104
  19. J. S. Yoo, S. J. Chang, S. H. Sun, Y. S. Hong, and S. S. Lee, J. Soc. Kor. Med. Diagnostics 17, 63 (2013).
  20. A. J. Baker and D. W. Pepper, Finite Elements 1-2-3, McGraw Hill, New York (1991).

Cited by

  1. Property of Focal Spot of Electron Beam Depending on the Anode Angle of X-ray Tube Using a Finite Element Method vol.25, pp.2, 2015, https://doi.org/10.4283/JKMS.2015.25.2.052